32
cod: ed. 09-2021
Frequency
reached
[Hz]
Head reached
[mbar]
I blowers
current
[A]
I centrifugal
separator
[A]
I input line
current
[A]
Condition
85 - 90
70 - 80
7,5 - 8
1,3 - 1,5
9,5 - 10
Suction inlet
open
48 - 52
330 - 338
8,5 - 8,8
1,3 - 1,5
7 - 7,5
Suction inlet
closed
To verify the correct operation of a variable
frequency suction unit, some dynamic type tests
can be carried out, such as the following.
With the machine switched on and suction inlet
not connected to the piping. The other test shall be
carried out with the suction inlet closed.
TURBO-SMART HP DIAGNOSTIC TEST
Fig. D
Summary of Contents for TURBO SMART HP
Page 1: ...WE LOVE WHAT WE DO English Instruction Manual TURBO SMART HP...
Page 2: ...2...
Page 30: ...30 cod ed 09 2021 SAME LEVEL INSTALLATION LEVEL BELOW INSTALLATION ASSEMBLY SCHEME...
Page 31: ...31 cod ed 09 2021 EN G LISH TURBO SMART HP CIRCUIT Fig C ELECTRICAL CONNECTIONS...
Page 33: ...33 cod ed 09 2021 EN G LISH EXAMPLES OF PIPING Fig E E TURBO SMART HP Fig F...
Page 34: ...34 cod ed 09 2021 Fig G TURBO SMART HP EXPLODED VIEW...
Page 35: ...35 cod ed 09 2021 EN G LISH Fig H CENTRIFUGAL SEPARATOR 023764 EXPLODED VIEW...
Page 36: ...36 cod ed 09 2021 Fig I UNI JET 75 SX 015663 EXPLODED VIEW...
Page 37: ...37 cod ed 09 2021 EN G LISH Fig L UNI JET 75 DX 015664 EXPLODED VIEW...
Page 38: ...38 cod ed 09 2021 Fig M AMALGAM SEPARATOR 043396 EXPLODED VIEW...
Page 39: ...39 cod ed 09 2021 EN G LISH Fig N CONTROL UNIT 023764 EXPLODED VIEW...
Page 40: ...40 cod ed 09 2021 Fig O TURBO SMART HP DIMENSIONS...
Page 41: ...41 cod ed 09 2021 EN G LISH Fig P TURBO SMART HP DIMENSIONS...
Page 42: ...42 cod ed 09 2021 TURBO SMART HP DIMENSIONS Fig Q...
Page 43: ...43 cod ed 09 2021 EN G LISH...