Run Times 

Apex Standard > 20 hours, Apex I.S > 12 hours, Apex Lite > 24hours (@ 3.5L/min via GFA 
filters). Microdust 

Pro

 with NiMH batteries typically 19 hours. Alkaline cells > 13 hours. (All run 

times may be reduced at low temperatures). 

Using the Dust Detective SAS Unit 

Installation 

Cutouts are provided in the foam for the Microdust 

Pro

 and Apex sampling units. Ensure both 

units are fully charged up or have new AA cells if required before insertion.  
 

Note: Ensure the Apex and Microdust pro are held firmly by the Velcro straps provided, 
otherwise damage to the instruments could occur. 
 

Connect the Apex pump to the sampling spigot (X). 
Connect the Microdust probe into the Adaptor Block and correctly align such that  the locating 
spigot on the PUF Filter Holder locks into the Probe Measuring Chamber. When locked correctly 
into position, the complete assembly is prevented from rotating or moving laterally on the probe. 
(see figure 4). The body of the probe will be located into the retaining clip. 

 

PUF Filter Adaptor 

For size-selective monitoring applications, it is necessary to load the appropriate foam filter(s) into 
the inlet adapter. If TSP (Total Suspended Particulate) monitoring is required, then no foams 
should be fitted within the inlet assembly.  
 
The type of foam filter loaded into the adapter determines the size of particulate matter being 
monitored by the Microdust 

pro

 and collected on the filter. Although the PUF foam inserts have 

been designed as size-selective filters to capture particles larger than a specified mean 
aerodynamic size, it is also possible to weigh the PUF inserts before and after measurement to 
establish both the total suspended particulate (TSP) value and the desired size fraction. 
 

Storage of PUF Filters 

The PUF filters should be kept in a clean and preferably air-conditioned environment.  
 

Inserting / Removing PUF Filters 

The PUF filters should be carefully inserted or removed from their respective cassettes using 
clean tweezers and plastic gloves. Avoid subjecting the filters to physical damage, creasing or 
folding. 
 
Filters should be loaded into the relevant housing as shown in Figure 3,4. 
 

Collection Filter Cassette 

A Filter Cassette is provided to accommodate a 37 or 25mm filter. The desired filter type (typically 
of the GFA type) should be conditioned and pre-weighed (if required) before being loaded into the 
cassette. Always handle filters with care to prevent contamination or physical damage. 
 

Summary of Contents for HB3341-01

Page 1: ...he copyright in this document which contains proprietary information is vested in CASELLA CEL Limited The contents of this document must not be used for purposes other than for which it has been supplied or reproduced or disclosed wholly or in part without the prior written permission of the author ...

Page 2: ...5 Collection Filter Cassette 5 Powering up and using the system 6 Checking the Instrument s Zero 6 Checking Probe Span 6 Sampling Pump 6 Starting a Sample Run 6 SERVICE 7 Enclosure Specification 8 Apex Pump Specification 8 Additional Apex Features 8 PUF FILTER SPARES 8 ORDERING INFORMATION 8 Figure 1 Components and controls for dust detective 9 Figure 2 External view of Dust Detective 9 Figure 3 P...

Page 3: ...he increase This need for real time particulate information is a requirement in general industry when looking at Total Suspended Particulate TSP and at inhalable and respirable dust levels as a health issue CASELLA CEL have introduced the Dust Detective Static Air Sampling SAS enclosure for exactly this application This accessory provides a simple solution to short to medium term fixed area monito...

Page 4: ... approaching dewpoint will be seen by the instrument as larger particles This biases results towards greater concentrations Also during these periods any air being drawn through the instrument which is high in humidity may condense on the optics and distort the light beam producing spurious results Any application requiring this type of monitoring regime will need a particulate monitor with intern...

Page 5: ...nlet adapter If TSP Total Suspended Particulate monitoring is required then no foams should be fitted within the inlet assembly The type of foam filter loaded into the adapter determines the size of particulate matter being monitored by the Microdust pro and collected on the filter Although the PUF foam inserts have been designed as size selective filters to capture particles larger than a specifi...

Page 6: ...o 25 cm H20 pressure drop Automatic flow control circuitry is used to maintain a stable sample flow rate over varying pressure drop conditions To switch the pump on press the enter key on the pump When running a real time flow meter symbol will appear on the display and a green flashing light will be visible at the top of the pump To switch the pump off press and hold the enter key and the pump wi...

Page 7: ... information please contact CASELLA CEL s Service Department at our Bedford premises We will be happy to provide quotations for individual repairs or provide annual maintenance under contract We recommend factory service by technicians trained and equipped to repair your instrumentation Should you wish factory repair assistance send your equipment in a package equivalent to the original packaging ...

Page 8: ...imensions 135 x 75 x 43 mm Additional Apex Features Elapsed time display Real Flow Rate indication Automatic flow fault indicator and shutdown Low battery warning indicator and bleeper Low battery shut down PUF Filter Spares Foam filter PM2 5 pk of 10 P118204 Foam filter PM10 pk of 10 P118206 Foam filter Respirable pk of 10 P118208 Pre weighed filters 25mm GFA P117000 Pre weighed filters 37mm GFA ...

Page 9: ...Components and controls for Dust Detective Accessory Pouch Filter Adapter Filter Conette Probe ON OFF Button Inlet Tube Zero Purge Hand Bellows Sampling Spigot x Apex Unit PM 2 5 Adaptor Microdust Unit Figure 1 Components and controls for dust detective Figure 2 External view of Dust Detective ...

Page 10: ...10 Figure 3 PUF Filter configurations Figure 4 Adaptor and Filter configurations ALTERATION WITOUT NOTICE The contents of this manual are subject to change without notice ...

Page 11: ...ng the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specifications are subject to change without notice The user is responsible for determining the suitability of this product ...

Reviews: