Powering up and using the system 

Checking the Instrument’s Zero 

Check the zero calibration for the Microdust Pro before every use. This can be achieved either as 
described in the Microdust 

Pro 

handbook, utilising the purge air bellows connected to the probe 

purge points. Alternatively to calibrate in situ, remove the inlet cowl, and connect the in-line air 
filter to the inlet tube, start the Apex pump. With clean air in the probe, the display should be 
zeroed. 
 

Checking Probe Span 

The span of the Microdust 

pro

 should be checked in accordance with the relevant User Manual. 

The span can also be checked by inserting the calibration filter and checking the reading. If 
required, adjust the span control according to the User Manual. 

 

Sampling Pump 

The Dust Detective enclosure is designed to use  the Apex range of personal samplers. This 
allows the user to take a standard personal sampling pump, charge the unit up and install inside 
the enclosure. The pumps are designed to provide a flow rate capability of up to 4.0 L/min whilst 
using a 37 mm GFA filter head (at up to 25 cm H

2

0 pressure drop). Automatic flow control 

circuitry is used to maintain a stable sample flow rate over varying pressure drop conditions.  
 
To switch the pump on, press the enter key on the pump. When running a real time flow meter 
symbol will appear on the display and a green flashing light will be visible at the top of the pump. 
To switch the pump off, press and hold the enter key and the pump will countdown to off. 
PUF filters are designed to operate at a flow rate of 3.5 L/min. 

 

Check the performance and calibrate the Apex sampling pump according to the manual and 
adjust the flow rate to 3.5 L/min. 

 

If the pump is unable to maintain the selected flow rate due to excessive pressure drop or an inlet 
blockage, the red LED will begin to flash and the warning tone will bleep rapidly.  

 

Starting a Sample Run 

1. 

Remove the protective dust cap and fit the sample inlet tube/head to the case (see 
fig 2). 

2. 

Insert the PUF filter foam(s) and in line sample filter (if required) (see figure 3,4). 

3. 

Switch the Microdust 

pro

 and Apex pump on. Confirm zero using the in-line filter 

connected to the sample inlet. 

4. 

Check that the Apex flow rate is set to 3.5L/min. 

5. 

Locate unit in required location. Start pump and Microdust data logger according to 
their respective user manuals. 

 

 

 
 
 

Summary of Contents for HB3341-01

Page 1: ...he copyright in this document which contains proprietary information is vested in CASELLA CEL Limited The contents of this document must not be used for purposes other than for which it has been supplied or reproduced or disclosed wholly or in part without the prior written permission of the author ...

Page 2: ...5 Collection Filter Cassette 5 Powering up and using the system 6 Checking the Instrument s Zero 6 Checking Probe Span 6 Sampling Pump 6 Starting a Sample Run 6 SERVICE 7 Enclosure Specification 8 Apex Pump Specification 8 Additional Apex Features 8 PUF FILTER SPARES 8 ORDERING INFORMATION 8 Figure 1 Components and controls for dust detective 9 Figure 2 External view of Dust Detective 9 Figure 3 P...

Page 3: ...he increase This need for real time particulate information is a requirement in general industry when looking at Total Suspended Particulate TSP and at inhalable and respirable dust levels as a health issue CASELLA CEL have introduced the Dust Detective Static Air Sampling SAS enclosure for exactly this application This accessory provides a simple solution to short to medium term fixed area monito...

Page 4: ... approaching dewpoint will be seen by the instrument as larger particles This biases results towards greater concentrations Also during these periods any air being drawn through the instrument which is high in humidity may condense on the optics and distort the light beam producing spurious results Any application requiring this type of monitoring regime will need a particulate monitor with intern...

Page 5: ...nlet adapter If TSP Total Suspended Particulate monitoring is required then no foams should be fitted within the inlet assembly The type of foam filter loaded into the adapter determines the size of particulate matter being monitored by the Microdust pro and collected on the filter Although the PUF foam inserts have been designed as size selective filters to capture particles larger than a specifi...

Page 6: ...o 25 cm H20 pressure drop Automatic flow control circuitry is used to maintain a stable sample flow rate over varying pressure drop conditions To switch the pump on press the enter key on the pump When running a real time flow meter symbol will appear on the display and a green flashing light will be visible at the top of the pump To switch the pump off press and hold the enter key and the pump wi...

Page 7: ... information please contact CASELLA CEL s Service Department at our Bedford premises We will be happy to provide quotations for individual repairs or provide annual maintenance under contract We recommend factory service by technicians trained and equipped to repair your instrumentation Should you wish factory repair assistance send your equipment in a package equivalent to the original packaging ...

Page 8: ...imensions 135 x 75 x 43 mm Additional Apex Features Elapsed time display Real Flow Rate indication Automatic flow fault indicator and shutdown Low battery warning indicator and bleeper Low battery shut down PUF Filter Spares Foam filter PM2 5 pk of 10 P118204 Foam filter PM10 pk of 10 P118206 Foam filter Respirable pk of 10 P118208 Pre weighed filters 25mm GFA P117000 Pre weighed filters 37mm GFA ...

Page 9: ...Components and controls for Dust Detective Accessory Pouch Filter Adapter Filter Conette Probe ON OFF Button Inlet Tube Zero Purge Hand Bellows Sampling Spigot x Apex Unit PM 2 5 Adaptor Microdust Unit Figure 1 Components and controls for dust detective Figure 2 External view of Dust Detective ...

Page 10: ...10 Figure 3 PUF Filter configurations Figure 4 Adaptor and Filter configurations ALTERATION WITOUT NOTICE The contents of this manual are subject to change without notice ...

Page 11: ...ng the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specifications are subject to change without notice The user is responsible for determining the suitability of this product ...

Reviews: