USER MAINTENANCE
End of procedure
Calibration Continued
8
Once the chamber has stabilized, compare the reference
temperature device and oven temperature display readings.
a.
If the readings are the same or the difference between the
two falls within the acceptable range of your protocol, the
display is accurately showing the chamber temperature.
The temperature calibration procedure is now complete
.
b.
See Step 9 if a difference falls outside the acceptable
range of your protocol again.
Reference Device
9
Repeat steps 2 – 8 up to two more times.
•
Three calibration attempts may be required to
successfully calibrate ovens that are more than ± 2°C out
of calibration.
Reference Device
If the temperature reading difference still falls outside your protocol after three (3) calibration
attempts, contact your distributor or