21
CAL 8
Function: Zero adjustment - calibration when any zeroing error occurs.
Used key
Display
Descriptions
2-CAL
Empty the load tray and press the setup key.
1234
The current weight value is displayed.
Confirm 'Stable' and press the setup key.
:Zero adjustment
- - -
Zero adjustment in progress...
Note 1. Use this function when zeroing is not passed for any shock to the load cell.
The range of zero adjustment is 0 ~ 2mV/V.
Note 2. Move to CAL-8 if the zero adjustment is over with no error.
Note 3. When zero point is too low, an error message "Err27" is displayed.
Note 4. When zero point is too high, an error message "Err 26" is displayed.
CAL 9
Function: Factor Calibration
Used key
Display
Descriptions
NOT USE
This function cannot be used because of
multi calibration.
FACtor
It means you entered the factor correction
mode.
:Save and
next Menu navigation
: Set value change
: Changing places
12345
The current factor is displayed.
Note 1. As this is a menu to set the weight setup with no weight, general users have no need to use it.
Note 2. This can be used only when the range of multi calibration in CAL 4-1 is set to 1.
"NOT USE" is displayed when the range of CAL 4-1 is set to 2 or larger.
Note 3. Enter a password to enter the factor correction mode.