26
5-3-1. General Function
F03
Function
Auto Power OFF
Display Meaning
F03. 00
Not used.
F03. 10
Automatic power off after 10 minutes in the waiting mode.
Setting range
(00 ~ 30)
F03. 30
Automatic power off after 30 minutes in the waiting mode.
Note 1. The power is automatically off if the defined time continues at the zero point after the automatic
power off is set.
F04
Function
Setting A/D Converting Speed
Display Meaning
F04. 10
10 rounds/second
F04. 20
20 rounds/second
Setting range
(00 ~ 99)
F04. 80
80 rounds/second
F05
Function
Setting digital filter
Display Meaning
F05. 10
Display of average for No. 10
F05. 30
Display of average for No. 30
Setting range
(00 ~ 50)
F05. 50
Display of average for No. 50
F06
Function
Setting vibration filter
Display Meaning
F06. 00
Vibration filter OFF
F06. 10
Compensation for the vibration value of 5 divisions (0.5d * 10)
Setting range
(00 ~ 99)
F06. 99
Compensation for the vibration value of 49.5 divisions (0.5d * 99)
Note 1. Apply this function to a place with heavy vibrations.
(The display response speed becomes slower when the vibration filter is applied.)
Note 2. This function should be adjusted appropriately to the site while the speed of weight variations in
F04 is being lowered little by little.
F07
Function
Setting Motion Detection Condition
Display Meaning
F07. 1
The 'Stable' lamp is lit if the weight changes within 0.5 division.
F07. 2
The 'Stable' lamp is lit if the weight changes within 1 division.
Setting range
(1 ~ 99)
F07. 10
The 'Stable' lamp is lit if the weight changes within 5 division.