6.3.4 CAL 4 (Hopper weight setting)
CAL 4-1
Function: Setting Multi Calibration Step
Range of set value: 2 ~ 5
Used key
Display
Descriptions
STEP- 2
Setting multi calibration for step 2
(for hopper system)
Using numeric keys
~
Set the range.
=Set,
=Cancel
STEP- 5
Setting multi calibration for step 5
(for hopper system)
CAL 4-2
Function: Setting Weight
Range of set value: 1 ~ 99,999
Used key
Display
Descriptions
LOAD 1
It means the weight setting mode.
(Number = multi calibration
number)
W=100.00
100.00 (unit: Kg or Ton)
Using numeric keys
~
Set the counterweight values
=Set,
=Cancel
W= 0.10
0.10 (unit: Kg or Ton)
Reference 1. Set the counterweight within 10~100% of the maximum capacity.
CAL 4-3
Function: Weight Calibration
Used key
Display
Descriptions
LoAd
Load the weight set in CAL 4-3 and press
the setup key.
12345
The current weight value is displayed.
Confirm 'Stable' and press the setup key.
=Set,
=Cancel
- - -
Span adjustment in progress...
Reference 1. Repeat 'CAL4-2' , 'CAL4-3' according to the number of STEPs set in 'CAL 4-1'.
At this instance the weight value should be calibrated to be greater than the previous values.
Reference 2. If the span calibration is completed without error, it will advance to CAL-4.
Summary of Contents for EXI-200AD
Page 1: ... 2 이미지 추가 ...
Page 20: ...21 4 External Explanation 4 1 External Specifications ...
Page 28: ...29 2 DCType ...
Page 29: ... 30 5 1 2 Load Cell Connector ...
Page 31: ... 32 5 1 3 RS 232 RS 485 Connector RS 485 RS 232 COM2 RS 232 COM1 ...
Page 98: ...99 MEMO ...
Page 99: ... 100 MEMO ...
Page 100: ...101 ...