37
6.3.1 CAL 1
(Maximum capacity calibration)
Function: Setting Maximum Value
Range of set value: 1 ~ 99,999
Used key
Display
Descriptions
C= 10000
Max value = 10000 kg
Using numeric keys
~
Enter maximum weight
=Set,
=Cancel
C= 10
Max value = 10 kg
Reference 1. The maximum capacity of the weighing scale can measure and display.
6.3.2 CAL 2 (Minimum division calibration)
Function: Minimum division and decimal position setting
Range of set value: 0.001 ~ 50
Used key
Display
Descriptions
d= 0.001
Minimum division 0.001 kg
d= 0.01
Minimum division 0.01 kg
d= 0.1
Minimum division 0.1 kg
d= 1
Minimum division 1 kg
Using numeric keys
~
Enter minimum division
=Set,
=Cancel
Push
when entering a decimal
d= 10
Minimum division 10 kg
Reference 1. When
CLEAR is pressed with decimal point calibrated, 'CAL2' is exited.
Reference 2. For external exploded view, set within the value of maximum capacity divided by the readability,
'1/30,000'. If the external exploded view is greater than '1/30,000', "Err20" will be displayed.
Reference 3. The decimal point positions is determined as the decimal position of the readability set from 'CAL2'.