39
F21
Function
Set the initial zero range
Display Meaning
F21. 02
Set the initial zero up to 2% of the maximum weight
F21. 10
Set the initial zero up to 10% of the maximum weight
Setting range
(02~20)
F21. 20
Set the initial zero up to 20% of the maximum weight
Note 1. Please consult with an engineer because setting 10 or larger value might affect the load cell greatly.
F23
Function
Setting the range of check for the excessive weight (weighing unit)
Display Meaning
F23 09
Excessive weight from the maximum 9 divisions
Setting range
(00~99)
F23. 99
Excessive weight from the maximum 99 divisions
F24(CI-201A)
Function
Backlight Operation
Display Meaning
F24 0
Backlight off
F24 1
Backlight on when any key is operated.
F24 2
Backlight on when the weight changes.
F24 3
Backlight on when it is 'Stable' after the weight changes.
F24 4
Backlight on when a key operates or the weight changes.
Setting range
(0~5)
F24 5
Backlight on all the time
Note. Although it is set to 5, press the power key shortly to turn off the backlight.