27
CAL 3-4
Function: Weight Calibration)
Used key
Display
Descriptions
LoAd
Load the weight set in CAL 4-3 and press the
setup key.
12345
The current weight value is displayed.
Confirm 'Stable' and press the setup key.
:Span adjustment
:End
- - -
Span adjustment in progress...
Note 1. CAL 3-3 and CAL 3-4 are repeated as many as STEP is set in CAL 3-1.
At this time, the weight value should be set to a value greater than the previous one.
Note 2. Move to CAL-1 if the span adjustment is over with no error.
Note 3. When zero point is low, an error message "ERR24" is displayed.
Note 4. When zero point is high, an error message "ERR25" is displayed.
Note 5. After finishing calibration, press the setup key for 2 seconds or more to convert to the weighing mode.
CAL 7
Function: Gravity Adjustment
Used key
Display
Descriptions
G-CAL
It means you accessed to the menu for the gravity
adjustment.
Gr-CAL
9.XXXX
Set the gravity for the production place.
:Save and next
Menu navigation
~ : Set value
change
:End
Gr-SET
9.XXXX
Set the gravity for the place to use the product.
Note 1. If the gravity of the indicator production place is different from that of the place to use, the gravity
adjustment can be done using this function.
Summary of Contents for CI-200 Series
Page 1: ...2 ...
Page 11: ...12 2 Specifications in Appearance 2 1 External Dimension CI 200A CI 201A ...
Page 12: ...13 CI 200S CI 200SC ...
Page 13: ...14 2 2 Front Panel Descriptions CI 200A CI 201A ...
Page 15: ...16 CI 200S CI 200SC ...
Page 78: ...79 MEMO ...
Page 79: ...80 MEMO ...
Page 80: ...81 MEMO ...
Page 81: ...82 ...
Page 82: ...83 ...