![CAS CI-100A Operation & Service Manual Download Page 6](http://html1.mh-extra.com/html/cas/ci-100a/ci-100a_operation-and-service-manual_2565919006.webp)
CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA
Corporation
www.cas-usa.com
4
1.2 Power Supply
:
1.2.1 4xAA size alkaline batteries: When the sign of battery is empty, this prompt you’d better to replace
batteries; When “Lo.bAt” displayed, this prompt you should replace batteries immediately, otherwise,
it will turn off automatically in 10s.
1.2.2 AC Adapter: 6-9V
DC,
≥
500mA, central positive:
1.2.3 USB power supply if USB interface is installed
1.2.4 Work current:
≤
15mA (with backlight off, no optional Bluetooth Module and no load cells)
≤
30mA (with backlight on, no optional Bluetooth Module and no load cells)
≤
50mA (with backlight on and one 350 load cell, no optional Bluetooth Module)
1.3 Display:
6-digit,7-segment , 1"(25mm) LCDs with 16 annunciators and blue backlight
1.4 Keypad:
6 push buttons
1.5 Environment:
1.5.1 Working temperature: -10°C to 40°C
1.5.2 Storage temperature: -20°C to 70°C
1.5.3 Humidity: 10 to 90% RH without condensation
1.6 Load cell Excitation:
1.6.1 Voltage: 5Vdc
1.6.2 Max. Current: 60mA (can power 4-350 ohm bridge, take care the limitation when USB is power
supply)
1.6.3 Signal connection: 4 or 6 lead with sense leads
1.6.4 Max Sensitivity: -3mV/V to +3mV/V
1.6.5 Load cell wiring (7 pin socket):
Pin1:
Red
Exci
Pin2:
White
Excitation -
Pin3:
Green
Pin4:
Blue
Signal -
Pin5:
Shield
Pin6:
Orange
Sense +
Pin7:
Black Sense -
If other special connection is used, please refer the additional paper
1.7 Communication:
1.7.1 Serial port1:
Full-duplex
RS232
1.7.2 Serial port2: USB (Virtual
RS232)
1.7.3 Optional
Serial port3: Bluetooth
1.7.3 Baud Rate:
Selectable: 1200-2400-4800-9600-19200-38400 bps
1.7.4 Data Output Format:
8N1, 7O1, 7E1, 7E2, 7O2
1.7.5 Protocol:
programmable