4.3 USER Submenu:
USER
Sub-
Menu1
Sub-
Menu2
Option
Default
Remark
Set
RESET
NO
NO
reset user parameters to default setting
YES
COM1 BAUD.RT 1200
9600
selection of com1's baud rate
2400
4800
9600
19200
38400
BYT.FMT
8N1
8N1
selection of com1's byte format:
(1)8N1=8 data bits, No parity check bit, 1 stop
bit; (2)7O1=7 data bits, 1 Odd parity check
bit, 1 stop bit; (3)7E1=7 data bits, 1 Even
parity check bit, 1 stop bit; (4)7O2=7 data
bits, 1 Odd parity check bit, 2 stop bit;
(5)7E2=7 data bits, 1 Even parity check bit, 2
stop bit;
7O1
7E1
7O2
7E2
OUT.MOD
NONE
PRTCMD
Selection com1 output mode: (1)NONE =No
communication; (2)CONT=continuously output;
(3)PRINT=output after PRINT key pressed;
(4)CMD=output after a request command is
received; (5)PRT.CMD= output after PRINT key
pressed or request command received;
(6)STABLE=output after scale is stable;
Note: use PRINT or CMD to output data, the
scale must be stable.
CONT
CMD
PRTCMD
STABLE
LAYOUT
MULTPL
MULTPL
com1 output content and format set:
(1)MULTPL= the following selected item in
OUT1
will be output using defined format;
(2)SINGLE= only displayed content and current
status will be output, it’s compatible with
NCI-SCP01; (3) EH-SCP= Command
–response mode, similar to Toledo PS60
protocol;
(
4
)
SCP-12 = only displayed content
and current status will be output, it’s compatible
with NCI-SCP12
(
NCI3835
)
;
SINGLE
EH-SCP
SCP-12
OUT1 SCAL.ID YES
NO
Yes/No=enable/disable output scale's ID
number, Prompt is “SCALE ID”
NO
GROSS
YES
NO
Yes/No=enable/disable output gross weight.
Prompt is “GROSS”
NO
TARE
YES
NO
Yes/No=enable/disable output tare weight.
Prompt is “TARE”
NO
NET
YES
YES
Yes/No=enable/disable output net weight.
Prompt is “NET”
NO