48
④
Press ENTER key in the above state. The [Zero calibration] starts automatically for
once.
CO
HC
%
ppm
CO
2
O
2
%
%
Lambda
AFR
NO
X
⑤
When the [Zero calibration] is completed, the values will be indicated on the
indication window as shown and the value in the first digit will blink.
CO
HC
%
ppm
CO
2
O
2
%
%
Lambda
AFR
NO
X
⑥
The
▲
key makes the blinking number increased. After the setup of the required
number, press
▼
key to move the blinking part to the next field. Use
▼
,
▲
keys to setup to standard gas value of gas cylinder, then press ENTER key. The each value of
CO, CO2 and O2 is entered the one displayed on the gas cylinder, but the HC value is entered the
N-Hexane value which is explained in 7-6. For example, HC value = the value is displayed on the
gas cylinder (propane HC value) x PEF value
.
If the HC value of the gas cylinder is 1002 ppm, multiply it by 0.5 of PEF value and the outcome
value of 501 ppm is gas calibration.