CS 8100SC Family User Guide (SM987)_Ed 01
43
Preparing and Positioning the Adult and Pediatric Patient
To prepare and position the patient, follow these steps:
1
Ask the patient to remove all metal objects.
2.
Ask the patient to wear a lead apron with a thyroid collar. Ensure that the apron lays flat across the
patient’s shoulders.
3.
Ask the patient to do the following:
•
Stand to the side and as far as possible from the cephalometric unit.
•
Open the hand fully and place it flat on the carpus panel.
IMPORTANT
Make sure that the:
•
hand, wrist, and forearm are in a
completely vertical position.
•
hand is approximately in the
middle of the carpus panel.
WARNING: To avoid exposure to other parts of the body, make sure
the patient stands as far away as possible from the X-ray emission
range and extends the hand to the carpus panel.