Disposal
P7000, P6000, TIVA, TCI & TIVA 99/106
1000SM00012 Issue 8
Information on Disposal for Users of Waste Electrical & Electronic Equipment
This
U
symbol on the product and/or accompanying documents means that used electrical and electronic products should not be mixed
with municipal waste.
If you wish to discard electrical and electronic equipment, please contact your CareFusion affiliate office or distributor for further
information.
Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human health and
the environment which could otherwise arise from inappropriate waste handling.
Information on Disposal in Countries outside the European Union
This symbol is only valid in the European Union. The product should be disposed of taking environmental factors into consideration.
To ensure no risk or hazard, remove the internal rechargeable battery and the Lithium battery from the control board and dispose of as
outlined by the local country regulations. All other components can be safely disposed of as per local regulations.
Disposal
Removal Procedure
1. Remove the six case retaining screws and washers located on the base of the pump.
2. Carefully separate the upper and lower case halves and disconnect cables.
3. Disconnect the battery cable from the Power Supply PCB.
4. Remove the two screws which secure the battery retaining plate.
5. Lift out the battery and retaining plate then disconnect the crimp terminals from the battery.
6. Detach the retaining plate from the battery.
Ensure the Pump is disconnected from the AC power supply and switched off before attempting to
service.
V
The Pump contains static-sensitive components and therefore strict ESD precautions should be
observed at all times.
Only use CareFusion recommended spare parts.
Following all spare part replacement and repair activities, testing must be performed in accordance
with the Performance Verification Procedure (PVP), see Chapter 3, Routine Maintenance.
Battery Removal
Battery
Screws (x2) /Washers (x2)
Summary of Contents for IVAC P6000
Page 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Page 87: ...Appendix C Fitting and Replacement Guidelines...
Page 98: ...Appendix E Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...
Page 105: ...Page Intentionally Left Blank...