Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 54/106
1000SM00012 Issue 8
Control PCB, Display PCB
Procedure
1. Disconnect the earth cable from the transmission assembly.
2. Remove the five PCB fixing screws and withdraw the Control PCB and Display PCB together.
3. Pull out the four standoffs/pins to separate the Control PCB and the Display PCB.
4. Reassemble in reverse order.
(
C
) Control PCB
(
B
) Display PCB
(
A
) EPROM
(
D
) PCB Screws (x5)
Earth cable
Grey ribbon cable
The transmission is shown removed for
clarity—it is not necessary to remove
the transmission assembly in order to
remove/refit the Control PCB and the
Display PCB.
Mandatory when serviced
Inspect and fit display spacers on pumps manufactured prior to July 1999 if the serial number is within the
ranges:
6001-00001 to 6001-13816
6002-00001 to 6002-11388
6003-00001 to 6003-01126
7001-00001 to 7001-13717
Procedure:
1) Visually inspect for the correct fit of the display pcb to the control pcb it is necessary to remove the control
pcb (with display pcb attached) from the upper case.
2) When removed, check that the display pcb is located on the four mounting spacers and is pressed as far as
possible onto the mounting spacer.
3) There shall be no gap between the spacer mounting face and either the control or display pcbs.
4) If a gap is present or the pcbs are not securely fitted the spacers are to be replaced with new spacer
(0000ME00387).
Summary of Contents for IVAC P6000
Page 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Page 87: ...Appendix C Fitting and Replacement Guidelines...
Page 98: ...Appendix E Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...
Page 105: ...Page Intentionally Left Blank...