Configured Options and Drug Protocol Records
P7000, P6000, TIVA, TCI & TIVA 94/106
1000SM00012 Issue 8
Enter the pump-specific information for your records on a copy of this page.
Syringes Enabled
Make
Size(s)
Make
Size(s)
Option
Range
Default
Setting
BACKOFF
Disabled/Enabled
Enabled
AC FAIL
Disabled/Enabled
Enabled
NEOI WARNING
2 - 10%
4%
EOI POINT
0.5 - 5.0%
1%
BEAM (Occlusion)
L0 - L7
L5
WEIGHT
0.1 - 150kg
70kg
PURGE RATE
100 - 500ml/h
500ml/h
HANDS FREE BOLUS
Disabled/Enabled
Enabled
DEFAULT BOLUS
0.1 - 25ml
1ml
NURSE CALL
Disabled/Enabled
Disabled
NURSE CALL INVERTED
Disabled/Enabled
Disabled
COMMS PUMP ADDRESS
1 -254
1
COMMS MONITOR ONLY
Disabled/Enabled
Enabled
COMMS ODD PARITY
Disabled/Enabled
Disabled
COMMS ASCII
Disabled/Enabled
Enabled
TCI
Disabled/Enabled
Enabled
DISPLAY TCI DECREMENT TIME
Disabled/Enabled
Enabled
Configured Options Record Sheet
IVAC® TCI & TIVA Syringe Pump
Software Version: V3R2
Hospital Name:
Serial No:
Configured by:
Date:
Software Version:
Approved by:
Date: