Configuration and Calibration
P7000, P6000, TIVA, TCI & TIVA 13/106
1000SM00012 Issue 8
The syringe pump software contains a number of configuration and test routines that can be accessed using a technical
access code. The majority of routines are 'menu' driven and are accessed by entering a technical access code as shown in
the table below.
Each menu (and certain individual options) has its own three-digit technical access code which is entered using the
following procedure:
1. Hold down the
H
button and turn the pump ON. The main display shows 000.
2. Enter the required access code “xxx” using the
M
keys in conjunction with the
NEXT
softkey (to move through
the digits).
3. When the required code is shown on the display, press the
OK
softkey to confirm.
Access Codes
Code
Title
Description
123
SELF TEST
Self test routine begins from the start. See Chapter 3, 'Routine Maintenance' for
further information
124
SELF TEST
Self test routine begins at power supply measurement
125
SELF TEST
Self test routine begins at declutch test
126
SELF TEST
Self test routine begins at pressure disc test
127
SELF TEST
Self test routine begins at beam test
128
SELF TEST
Self test routine begins at nurse call test
167
COMMS LEARN MODE
Configuration set via comms interface
168
COMMS TEACH MODE
Configuration output to another device (pump)
176
LINEAR DIAGNOSTIC
Linear potentiometer test
222
SELF TEST
Self test routine begins at display test
223
SELF TEST
Self test routine begins at keypad button test
243
SYRINGE SIZE CALIBRATION
Syringe size measurement calibration. See 'Calibration Procedures' on the following
pages for instructions
251
CONFIGURATION MENU
Configuration of drugs, options, syringes and real-time clock. See 'Configuration
Options (251)' on the following pages for further details
253
LINEAR POT CALIBRATION
Linear potentiometer calibration. See 'Calibration Procedures' on the following
pages for instructions
263
PRESSURE CALIBRATION
Line pressure calibration. See 'Calibration Procedures' on the following pages for
instructions
359
LANGUAGE
Configure language
376
SERVICE ACCESS MENU
Review service log, errors and hours of use. See 'Service Access Options' in Chapter
3 for further details
611
TOTAL MEMORY CLEAR
Reset memory. Note that the pump will require full calibration
717
BEAM OCCLUSION CALIBRATION
Beam occlusion calibration. See 'Calibration Procedures' on the following pages for
instructions
901
DEDICATION MODE
Set pump to operate in fully dedicated mode. Model P7000 only, see 'Dedicated
Options (901/902)' below for further details
902
DEDICATION MODE
Set pump to semi-dedicated mode. Model P7000 only, see 'Dedicated Options
(901/902)' below for further details
Entering an Access Code
Summary of Contents for IVAC P6000
Page 1: ...s IVAC Syringe Pumps Models P7000 P6000 TIVA TCI TIVA Technical Service Manual...
Page 87: ...Appendix C Fitting and Replacement Guidelines...
Page 98: ...Appendix E Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...
Page 105: ...Page Intentionally Left Blank...