Spare Parts Replacement Procedures
P7000, P6000, TIVA, TCI & TIVA 49/106
1000SM00012 Issue 8
RS232 Connector
Procedure
1. Remove the four screws which secure the cover plate to the underside of the lower case. Hinge the cover plate open
and remove it.
2. Disconnect the RS232 ribbon cable from the Control PCB.
3. Remove the two screws which secure the RS232 connector to the lower case and slide out the connector assembly
and gasket.
4. Reassemble in reverse order.
Refitting notes:
1) Apply loctite (243) to the threads of the RS232 screws (D) before assembly.
Spare Parts
Item
Description
Part Number
A
CORD SEALING SILICONE ID 0.95
1000ME01087
B
PLATE BASE P SERIES
6000ME00026
C
SCREW M3X8 CSK HD POSI SS
0000ME00268
D
SCREW MALE/FEMALE M3.4 BRASS & NICKEL
0000ME00279
E
ASSY RS232/NC CONNECTOR INSULATED
6000SP00024
F
GASKET RS232 MOULDED & CAP A4
1000ME01106
(
F
) Gasket
(
E
) RS232 connector
(
D
) Screws (x2) /
Loctite 243
(
C
) Cover plate screws (x4)
(
B
) Base plate
(
A
) Cord sealing