Alaris™ VP Plus Guardrails™ Volumetric Pump
Main Display Features
1000DF00660 Issue 9
8/48
Main Display Features
Main Display
Infusion Status/Drug Name/
Profile Name/Primary or
Secondary
Infusion Rate
Dose Rate
Volume to be Infused
Volume Infused
Time remaining
Softkeys
Softkey Identifiers
Clear Volume Infused
Set VTBI Option
NORADrenaline
RATE
VOLUME
VTBI
25
.0
ml/h
mcg/kg/min
1
.2
1 h 48 m 00 s
VTBI
VOLUME
ml
ml
45
.0
50
.0
mmHg
28
60
Line Pressure
Downstream Occlusion
Alarm Limit
Screen Icons
Symbol
Description
Time
icon - Indicates time remaining before VTBI will be completed . If the time is greater than 24 hours then 24+
will be displayed . Also indicates pressure sample time on the Pressure Trend screen .
N
Battery
icon - Indicates battery charge level to highlight when the battery will require recharging .
Line Pressure
icon - Shows the current in-line pressure value in mmHg .
Downstream Occlusion Alarm Limit
icon - Shows the downstream occlusion alarm limit in mmHg .
?
Indicates that the value entered is outside of the Guardrails soft limits . When the warning is overridden this
indicates Guardrails safety protocols are in use .
!
Indicates that the value entered is outside of the Guardrails hard limits . This warning cannot be overridden . This
symbol is also used to prompt the user to set the rate .
Indicates that the Pump is running at a rate below a Guardrails soft limit .
Indicates that the Pump is running at a rate above a Guardrails soft limit .