Alaris™ VP Plus Guardrails™ Volumetric Pump
Controls and Indicators
1000DF00660 Issue 9
6/48
Controls and Indicators
Controls
Symbol
Description
a
ON/OFF
button - Press once to switch the Pump on . Press and hold down for approximately three
seconds to switch the Pump off .
b
RUN
button - Press to start the infusion . The green LED will flash during infusion .
h
HOLD
button - Press to put the infusion on hold . The amber LED will be lit while on hold .
c
MUTE
button - Press to silence active alarm, warning or prompt for approximately 2 minutes . The
audio from the active alarm, warning or prompt will continue to sound every 2 minutes until the
condition has been resolved . To silence the active audio for an extended interval of 15 minutes; press
CANCEL
, then press and hold the
MUTE
button until 3 rapid beeps are heard . Once the 15 minute
interval has timed out the mute interval returns to 2 minutes .
i
PRIME/BOLUS
button - Press to access
PRIME
or
BOLUS
softkey . Press and hold down softkey to
operate .
PRIME - primes the infusion set with fluid when setting up an infusion for the first time .
• Pump is on hold .
• Infusion set is not connected to a patient .
• Volume infused (VI) is not added to the total volume infused displayed .
BOLUS - delivers fluid or drug at an accelerated rate .
• Pump is infusing .
• Infusion set is connected to patient .
• Volume infused (VI) is added to the total volume infused displayed .
d
OPTION
button - Press to access optional features .
e
PRESSURE
button - Use this button to display the line pressure, trend display, and adjust the
downstream occlusion alarm limit .
f
CHEVRON
keys - Double or single for faster/slower increase/decrease of values shown on display .
g
BLANK SOFTKEYS
- Use in conjunction with the prompts shown on the display .
Indicators
Symbol
Description
S
AC POWER
indicator - When illuminated the Pump is connected to an AC power supply and the
battery is being charged .
j
BATTERY
indicator - When illuminated the Pump is running on the internal battery . When flashing,
the remaining battery run time is limited .