Alaris™ VP Plus Guardrails™ Volumetric Pump
Prompts
1000DF00660 Issue 9
32/48
Prompts
Prompts alert the user but may not stop the infusion and are indicated by an audible sound, a flashing amber alert indicator and a
message on the display .
1 . Check the display for a prompt message . Press
c
to silence the sound for 2 minutes,
CANCEL
to clear the message .
2 . Rectify the cause of the prompt or proceed with caution .
Display
Cause
Action
ATTENTION
Pump left on hold for 2 minutes without starting
the infusion .
• Review Pump setup .
• Start infusion or turn off Pump .
SET VTBI
No VTBI/flow sensor .
• Set VTBI or fit flow sensor .
SET NOT FITTED
No infusion set fitted .
• Fit infusion set .
LOCKED
Rate change attempted while locked .
• Unlock rate to adjust infusion settings .
ADD DRUG
Drug selection required .
• Press
d
to access options menu .
• Select
DRUGS A-Z
using the
f
keys .
• Press
OK
to confirm .