1000DF00741 Issue 5
35/52
Alaris™ PK Plus Syringe Pump
Specifications
Alarm Conditions
Drive Disengaged
Occlusion
Attention (Nurse Callback)
Check Syringe
Near End Of Infusion
End of Infusion
Battery Low
Battery Empty
AC Power Fail
Titration not confirmed
Internal Malfunction
Concentration not Permitted
Dose Would Exceed
Target Would Exceed
Dose Under
Dose not Permitted
Bolus Dose not Permitted
Rate not Permitted
Bolus Dose Under
Bolus Dose Over
Weight Outside Limit
Environmental Specifications
Operating Temperature
0°C - +40°C
Operating Relative Humidity
20% - 90%
Operating Atmospheric Pressure
700hPa - 1060hPa
Transport and Storage Temperature
-30°C - +50°C
Transport and Storage Relative Humidity
10% - 95%
Transport and Storage Atmospheric Pressure 500hPa - 1060hPa
Electrical/Mechanical Safety
Complies with EN/IEC60601-1 and EN/IEC60601-2-24.
Potential Equalisation Conductor
The function of the Potential Equalisation Connector (Conductor) is to provide a direct connection between the Pump and the potential
equalisation busbar of the electrical installation. To use the Potential Equalisation Connector, connect the Potential Equalisation
Connector on the Pump to the potential equalisation busbar of the electrical installation.
EMC
Complies with EN/IEC60601-1-2, EN/IEC60601-2-24