![CareFusion Alaris PK Plus MK4 8005PK201 Directions For Use Manual Download Page 26](http://html1.mh-extra.com/html/carefusion/alaris-pk-plus-mk4-8005pk201/alaris-pk-plus-mk4-8005pk201_directions-for-use-manual_2557213026.webp)
1000DF00741 Issue 5
25/52
Alaris™ PK Plus Syringe Pump
Operations During Use
Operations During Use
End of Operation
This option will only appear in the options menu when the infusion has been stopped.
1. Press the
d
button to access the options menu.
2. Select the
END OF OPERATION
option using the
f
keys.
3. Press the
OK
softkey indicated on the screen.
Note:
Selecting this option will reset parameters for a new patient.
TCI Mode
When the Pump is on hold in predictive TIVA mode, the user is able to switch from TIVA to TCI mode.
1. Press the
d
button to access the options menu.
2. Using the
f
keys, select the
TCI MODE
.
3. Press the
OK
softkey indicated on the screen. A confirmation screen will be displayed.
Note:
When the mode is changed to TCI mode, the initial target will be set to zero.
TIVA Mode
When the Pump is on hold in TCI mode, the user is able to switch from TCI to predictive TIVA mode.
1. Press the
d
button to access the options menu.
2. Using the
f
keys, select the
TIVA MODE
.
3. Press the
OK
softkey indicated on the screen. A confirmation screen will be displayed.
Note:
When the mode is changed to predictive TIVA mode, the initial doserate will be set to zero.
DECREMENT CONC.
In TCI mode only:
1. Press the
d
button to access the options menu.
2. Select
DECREMENT CONC
.
3. Select the required
DECREMENT CONC
and press the
OK
softkey to exit.
Trend Size
The user is able to select the Trend Size of the Concentration Prediction graph.
1. Press the
d
button to access the options menu.
2. Using the
f
keys, select
TREND SIZE
.
3. Using the
f
keys, select the required
TREND SIZE
option (
5 Mins
,
15 Mins
,
30 Mins
or
60 Mins
).
4. Press the
SELECT
softkey indicated on the screen.
5. Press the
RESIZE
softkey to rescale the vertical axis of the graph. The initial displays calculates the scale so the peak value fills graph.
If the trend is downward the graph only fills lower part and the
RESIZE
option forces it to rescale.
Text/Graph Display
When in TCI mode, the user is able to select a numerical or graphical display.
1. Press the
d
button to access the options menu.
2. Using the
f
keys, select the display mode (
TEXT
or
GRAPH DISPLAY
). The options menu shows the available display mode
option.
3. Press the
OK
softkey indicated on the screen.
Dosing Summary
1. Press the
d
button to access the options menu.
2. Select the
DOSING SUMMARY
option using the
f
keys and press the
OK
softkey.
3. Press the
QUIT
softkey to exit the menu.