Alaris™ PK Plus Syringe Pump
Prompts
Prompts
Display
Icon
Description and Troubleshooting Guide
DOSE WOULD EXCEED
The infusion rate has been set to a value which exceeds a Soft Alert. Check infusion
setting, to continue with infusion at set rate press the
b
button and then confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press
the
NO
softkey and adjust rate below Soft Alert.
DOSE UNDER
The infusion rate has been set to a value which is under a Soft Alert. Check infusion
setting, to continue with infusion at set rate press the
b
button and then confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press
the
NO
softkey and adjust rate above Soft Alert.
DOSE NOT PERMITTED
The infusion rate has been set above a Hard Limit. Check infusion setting and adjust rate
to appropriate required rate.
TARGET WOULD EXCEED
The target has been set to a value which exceeds a Soft Alert. Check infusion setting,
to continue with infusion at set target press the
b
button and then confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press
the
NO
softkey and adjust rate below Soft Alert.
BOLUS DOSE OVER
The bolus dose has been set to a value which exceeds a Soft Alert. Check the
bolus setting, to continue with the bolus press the
b
button and then confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press
the
NO
softkey and adjust dose below Soft Alert.
BOLUS DOSE UNDER
The bolus dose has been set to a value which is under a Soft Alert. Check the
bolus setting, to continue with the bolus press the
b
button and then confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press
the
NO
softkey and adjust dose above Soft Alert.
BOLUS DOSE NOT PERMITTED
The bolus dose has been set above a Hard Limit. Check bolus setting and adjust to
appropriate required dose.
WEIGHT OUTSIDE LIMIT
The patient weight has been set to a value which exceeds or is under a Soft Alert. Check
the weight setting, to continue press the
b
button and then confirm
OVERRIDE LIMIT
by pressing the
YES
softkey. If
OVERRIDE LIMIT
is not required press the
NO
softkey and
adjust the value within the limits.
RATE NOT PERMITTED
The infusion rate has been set above a Hard Limit. Check infusion setting and adjust to