1000DF00741 Issue 5
29/52
Alaris™ PK Plus Syringe Pump
Configured Options
Configured Options
This section comprises of a list of options which are configurable. Some can be entered via the Pump configuration menu (available in
Technician Mode) and others through the Alaris PK Editor Software.
Enter the access code on Alaris PK Plus Syringe Pump for Configured Options, see the Technical Service Manual for details.
w
Access codes should only be entered by qualified technical personnel.
Use Alaris PK Editor to configure general options, drug library and units enabled for each profile and to configure Syringe Brands and
Models to be enabled.
Clock Set
1. Select
CLOCK SET
from the Configured Options menu using the
f
keys and press the
OK
softkey.
2. Use the
f
keys to adjust the date displayed, pressing the
NEXT
softkey to access the next field.
3. When the correct time and date are displayed press the
OK
softkey to return to the Configured Options menu.
Language
This option is used to set the language of messages shown on the Pump display.
1. Select
LANGUAGE
from the Configured Options menu using the
f
keys and press the
OK
softkey.
2. Use the
f
keys to select the language.
3. When the desired language has been selected press
SELECT
softkey to return to the Configured Options menu.
Contrast
This option is used to set the contrast on the Pump display.
1. Select
CONTRAST
from the Configured Options menu using the
f
keys and press the
OK
softkey.
2. Use the
f
keys to select a contrast ratio value. The contrast of the display will change when scrolling through the numbers.
3. When the desired value has been reached press the
OK
softkey to return to the Configured Options menu.
Alaris PK Plus Syringe Pump General Options
1. Select
GENERAL OPTIONS
from the Configured Options menu using the
f
keys and press the
OK
softkey.
2. Select the option required to enable/disable or adjust and press the
MODIFY
softkey.
3. When all the required modifications have been carried out press the
QUIT
softkey.
4. Either select the next configuration option from the menu or turn the Pump off, returning it to operation as required.
NURSE CALL FITTED
Enables Nurse Call (hardware option).
NURSE CALL INVERT
When enabled, the nurse call output is inverted.
RS232 SELECTED
Sets the Pump's communications to use RS232 (hardware option). The
NURSE CALL FITTED
option must
be enabled to allow RS232 to function.
Power Down Sequence
Enter the access code on Alaris PK Plus Syringe Pump for alternative Power Down Sequence, see the Technical Service Manual for details.
w
Access codes should only be entered by qualified technical personnel.
ENABLED
When running TCI or TIVA with predictive TCI the Pump may only be powered down by stopping the infusion,
selecting NEW OPERATION from the options menu, confirming the selection and then power down the Pump.
DISABLED
In TCI or TIVA with predictive TCI the Pump may be powered down after putting the Pump on hold.