4.1

 

Function 

The mattress system independently and without 

manual adjustment controls the inner pressure 

of the mattress according to the weight, height, 

and position of the user. No manual action needs 

to be performed to adjust the inner pressure of 

the mattress. This function works as follows:  

 

1. At start-up, automatic setting of the inner 

pressure of the mattress is always carried out 

according to the weight and height of the user. 

2. If the user moves noticeably or changes their 

position, the system will independently control 

the inner pressure of the mattress. 

3. The system performs an automatic setting at 

fixed intervals even if no significant changes have 

occurred. 

After automatic setting of the mattress inner 

pressure, the system returns to the previously 

selected program. At start-up, Pulsating Mode is 

always pre-set. 

4.2

 

Programs 

There are three programs to choose from: 

 

CLP

 means that the air pressure in all of the air 

cells is the same in the whole mattress. 

 

Alternating mode

 means that the air pressure in 

the air cells are different and alternates regularly 

after chosen cycle periods.  

 

Pulsating mode

 combinates CLP with the 

alternating function.  

 

Choose program by pushing the button for the 

program. We recommend the Pulsating mode 

which is also the preset mode.  

 

 

1.

 

Constant low pressure mode (CLP)

 

No cycle period is needed. 

 

 

 

2. Alternating mode.

 The cycle period 

can be changed according to the user 

needs and requirements. Choose 

between 10, 15, 20 or 25 minutes. The longer 

the cycle period, the slower the alternations. 

We recommend a basic setting of 10 minutes. 

 

 

 

 

 

 

 

3. Pulsating mode. 

The cycle period can 

be changed according to the user needs 

and requirements. Choose between 10, 

15, 20 or 25 minutes. The longer the cycle 

period, the slower the alternations. We 

recommend a basic setting of 10 minutes. 

4.3

 

Maximum pressure (caring mode) 

With this function, the entire mattress 

is inflated and provides firmed 

support. This function reverts to the 

previous setting after 20 minutes. The 

function should be used when caring the user, 

shifting the user’s position or moving the user in 

or out of bed. 

4.4

 

Panel lock 

Press the Panel Lock button to lock or 

unlock the control panel. The button 

indicates when the panel has been 

locked. The screen locks automatically if left 

untouched for five minutes. This is to prevent 

the settings being changed accidentally. 

To unlock, press the button for 2 seconds. 

4.5

 

Comfort settings 

The pressure can be 

increased in two steps 

depending on the user’s 

comfort requirements. This increase is made 

based on the automatic setting in 5.1. 

 

The selected setting is shown by a green light. 

 

 

 

 

 

 

 

 

 

Note! 

When only parts of the mattress are under 

load, for example, in the case of amputees  

– it may be necessary to raise the setting using 

the comfort settings. 

Summary of Contents for CuroCell A4 Cirrus

Page 1: ...EN Instructions for use 2022 06 27 CuroCell A4 IQ Cirrus Air mattress systems Instructions for use item number 95 001455 EN0000...

Page 2: ...9 Transport function 9 4 10 Pack Go function 9 4 11 Restart 9 4 12 Power failure 9 4 13 Notifications 9 4 14 Maximum pressure notifications 9 4 15 Table of notifications 9 5 Operation CuroCell IQ 11...

Page 3: ...duct functionality by performing a hand check 4 When the product is used for individuals needing special supervision such as children continuous monitoring is required 5 The mattress is protected by a...

Page 4: ...epair 15 Never connect anything other than the Care of Sweden supplied power supply to the control unit power cable connector 16 Never use the external communication input 3 5mm connector this input s...

Page 5: ...ystem consists of a control unit and a mattress and is intended to be used for prevention and as an aid in the treatment of pressure ulcers pressure injuries PU PI 1 3 Intended user The mattress syste...

Page 6: ...it running over it with the bed wheels or getting it jammed when raising or lowering the bed 4 Open the lid on the air tube connector marked CPR and connect it to the side of the control unit 5 A clic...

Page 7: ...wer supply from the wall socket and leave the control unit hanging on the bed during transport The mattress will retain air for at least 12 hours We recommend using this function for short periods onl...

Page 8: ...figure 1 2 The mattress starts to inflate This takes about 20 minutes depending on the size of the mattress While the mattress is inflating the mute the information signal and information signal diod...

Page 9: ...on for deflating the system Alternating mode Pulsating mode CLP mode Constant low pressure Maximum pressure caring mode Information signal Incorrect connection of the air connector CPR Cycle time sett...

Page 10: ...ode 1 Constant low pressure mode CLP No cycle period is needed 2 Alternating mode The cycle period can be changed according to the user needs and requirements Choose between 10 15 20 or 25 minutes The...

Page 11: ...ete Carefully fold the mattress in half place the control unit between the folds of the mattress and place the system in a transportation bag accessory or equivalent for protective storage Ensure that...

Page 12: ...e automatic setting has been restarted too many times during the automatic setting period 9 The mattress control parameters have not been read Connect the CPR 10 The mattress control parameters have b...

Page 13: ...osition of the user This takes about 20 minutes During this time the diode above the check symbol flashes When the diode stops flashing the mattress is set and the inner pressure has adjusted accordin...

Page 14: ...carried out according to the weight and length of the user 2 If the user moves noticeably or changes their position the system will independently control the inner pressure of the mattress 3 The syst...

Page 15: ...monary resuscitation See 3 2 5 6 Transport function See 3 3 5 7 Pack Go The product may be packed as follows Ensure that no one is lying on the mattress On the control panel press the lock unlock butt...

Page 16: ...problem remains contact the technical support 5 Automatic setting failure The correct pressure has not been reached within the time limit If the problem remains contact the technical support 6 Leakag...

Page 17: ...in position This means that no manual action needs to be performed to adjust the inner pressure of the mattress to conform to the user With CuroCell A4 there are three programs to choose from CuroCell...

Page 18: ...16 1 6 2 Mattress 1 Mattress 2 Hygiene cover top part 3 Air cells 4 Foam Core 5 Inner cover 6 Tubing set 7 CPR quick deflation 8 Hygiene cover bottom part 2 3 4 5 6 7 8...

Page 19: ...xide max 1 5 Excrement and blood stains must be removed as soon as possible using cold water Carefully follow local instructions and the instructions for the detergent Multi layer covers should be div...

Page 20: ...ed and inspected to maintain functionality and performance Service and maintenance must always be performed by Care of Sweden or one of its authorized technicians Only use spare parts approved by Care...

Page 21: ...oves around Check that the mattress is fastened to the bed frame with the straps underneath two at the head end and two on each of the long sides Some air cells have less air This is normal with a Pul...

Page 22: ...Use only power supply P N WR9QE1500LRPCIMG3138 Efficiency level VI Power consumption Max 18W Electrical classification Class II Type BF Fuse No Fuse Mode of operation CuroCell A4 Constant Low Pressure...

Page 23: ...ical Device Regulation EU 2017 745 Manufacturer Medical Device Distributor UDI XXXX XX XX Year Month Day Foot placement User information category Recommended user weight Counteracts shear Do not rotat...

Page 24: ...disassembled and are sorted as electronic waste Mattress A used CuroCell mattress should be taken to a recycling center The product is sorted as combustible waste 12 3 Returns Contact Care of Sweden...

Page 25: ...reofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns v g 4A SE 514 33 Tranemo SWEDEN Distributed by C...

Reviews: