CuroCell® A4 Cirrus
6
Note!
•
During the automatic setting, try to avoid
larger movements on the mattress. Otherwise,
you will get a notification on that the desired
value could not be reached within the time
limit and the weighing must start over.
•
The mattress must be inflated before the user
can lie on the mattress.
•
Once automatic setting is complete, the
control unit switches to a basic setting of
Pulsating mode (when used for the first time)
or to the previous setting.
4
Operation CuroCell® A4 Cirrus
Operation CuroCell® A4 Cirrus (how to get started)
1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1.
2. The mattress starts to inflate. This takes about 20 minutes depending on the size of the mattress.
While the mattress is inflating, the ”mute the information signal” and ”information signal” diodes
light up in orange. When these diodes goes out, the user can be placed on the mattress.
3. The pulsating mode setting is pre-set. If a different setting is desired, it can be selected when the
user is placed on the mattress.
4. The control unit sets the inner pressure of the mattress according to the weight, length, and
position of the user. This takes about 20 minutes. During this time, the diode above the selected
program flashes. When the diode stops flashing, the mattress is set and the inner pressure has
adjusted according to the user.
5. Perform a hand check to ensure that the settings are correct.
Figure 1. On/Off switch.