CuroCell® IQ Cirrus
12
5.1
Function
The mattress system independently and without manual adjustment controls the inner pressure of
the mattress according to the weight, length, and position of the user. No manual action needs to
be performed to adjust the inner pressure of the mattress. This function works as follows:
1. At start-up, automatic setting of the inner pressure of the mattress is always carried out
according to the weight and length of the user.
2. If the user moves noticeably or changes their position, the system will independently control
the inner pressure of the mattress.
3. The system performs an automatic setting at fixed intervals even if no significant changes have
occurred.
After automatic setting, the system returns to Pulsating Mode.
Button
Function
Mute the information signal
Pack & Go
®
. Function for deflating the system
Maximum pressure (caring mode)
Information signal
Incorrect connection of the air connector (CPR)
Check-symbol. System is ready to use
Notification diodes