3-6
70-02006-01 A
Powerheart® AED G3 9300A and 9300E
How to Perform a Rescue
Step 4: Analyze the heart rhythm
When the AED prompts…
Do this ...
“ Do Not Touch Patient. Analyzing Do not touch the patient.
heart rhythm.”
Wait for the next prompt.
The AED begins analyzing the
cardiac rhythm of the patient.
During the analysis phase, you may hear one or more of these prompts:
If the AED prompts… This is the problem…
Do this ...
“Check pads.”
The pads are disconnected
Ensure that the connector
from the AED.
is properly plugged into
Or...
the AED.
The pads are not properly
placed or are loose.
“Analysis interrupted.
The patient is excessively
Remove the electronic
Stop patient motion.”
jostled or there is a strong
device or stop the excessive
The AED restarts the
electromagnetic emitting
motion.
analysis.
device nearby (within 5
meters).