CARDIAC SCIENCE AEDs
INSTRUCTIONS FOR USE
4
INTELLISENSE
®
BATTERY
INSTALLATION
•
Insert battery as shown.
•
Push firmly to snap into place.
•
Open the lid for 5 seconds.
•
The
STATUS INDICATOR
turns
GREEN
(Does not apply to 9300C).
ABOUT THE INTELLISENSE BATTERY
•
When the last battery indicator (LED) is red, the battery is low. Replace the battery right away.
•
A new battery typically takes 10 seconds to charge the AED to maximum energy.
•
Output voltage: 12VDC (max)
•
Batteries are non-rechargeable
•
Lithium contents: 9.2g (max)
•
Check local regulations for disposal information
MODEL
FULL OPERATIONAL REPLACEMENT
GUARANTEE (FROM DATE OF INSTALLATION)
TYPICAL SHOCKS
9146 Lithium
9143 Lithium
4 Years
1 Year
up to 290 shocks
up to 100 shocks
Additionally, batteries are warranted against defects in materials and workmanship for 7 years from date of purchase.