ENERGY LEVELS
Table A1 - Ultra-Low Variable Energy (150 VE) AED Models 9300 Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(ms)
Voltage*
(Volts)
Duration*
(ms)
Energy**
(Joules)
25 1393
3.3
743
3.2
145-196
50 1420
4.5
909
3.2
128-173
75 1430
5.8
973
3.2
116-156
100
1434 7.0 1007 3.2
108-146
125
1437 8.3 1027 3.2
102-138
150
1439 9.5 1040 3.2
98-132
175 1441
10.8
1049
3.2 95-128