CARDIAC SCIENCE AEDs
INSTRUCTIONS FOR USE
22
Table A2 - Low Variable Energy (200 VE) AED Models 9300 Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(ms)
Voltage*
(Volts)
Duration*
(ms)
Energy**
(Joules)
25 1609
3.3
858
3.2
193-260
50
1640 4.5 1050 3.2
170-230
75
1651 5.8 1124 3.2
155-209
100
1656 7.0 1163 3.2
144-194
125
1660 8.3 1186 3.2
136-184
150
1662 9.5 1201 3.2
131-176
175 1663
10.8
1212
3.2 126-170
Summary of Contents for FirstSave AED 3g