CARDIAC SCIENCE AEDs
INSTRUCTIONS FOR USE
23
Table A3 - High Variable Energy (300 VE) AED Models 9300 Waveform
* All values are typical.
** Allowable energy range.
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(ms)
Voltage*
(Volts)
Duration*
(ms)
Energy**
(Joules)
25 1869
3.3
997
3.2
260-351
50
1906 4.5 1220 3.2
230-311
75
1918 5.8 1306 3.2
210-283
100
1925 7.0 1351 3.2
195-263
125
1928 8.3 1378 3.2
184-248
150
1931 9.5 1396 3.2
176-238
175 1933
10.8
1408
3.2 170-230
Summary of Contents for FirstSave AED 3g
Page 27: ...P N 112 0036 401 Rev A...