CAUTION:
Temperature Extremes
Exposing the AED to extreme environmental conditions outside of its operating parameters may compromise
the ability of the AED to function properly. The RescueReady® daily self-test verifies the impact of extreme
environmental conditions on the AED. If the daily self-test determines environmental conditions outside of the
AEDs operating parameters, a “SERVICE REQUIRED” alert will be issued to prompt the user to move the
AED to environmental conditions within the acceptable operating parameters at once.
CAUTION:
Lithium Sulfur Dioxide Battery
Pressurized contents: Never recharge, short circuit, puncture, deform, or expose to temperatures above 65°C
(149°F). Remove the battery when discharged.
CAUTION:
Battery Disposal
Recycle or dispose of the lithium battery in accordance with all federal, state and local laws. To avoid fire and
explosion hazard, do not burn or incinerate the battery.
CAUTION:
Use only Manufacturer Approved Equipment
Using batteries, pads, cables, or optional equipment other than those approved by the manufacturer may
cause the AED to function improperly during a rescue.
CAUTION:
Possible Improper AED Performance
Using pads that are damaged or expired may result in improper AED performance.
CAUTION:
Serial Communication Cable
The AED will not function during a rescue when the serial communication cable is connected to its serial port.
When the serial communication cable is connected to the AED during a rescue, the prompt
“Remove Cable
to Continue Rescue”
will be heard until you remove the serial communication cable.
CAUTION:
Possible Radio Frequency (RF) Susceptibility
RF susceptibility from cellular phones, CB radios and FM 2-way radio may cause incorrect rhythm recognition
and subsequent shock advisory. When attempting a rescue using the AED, do not operate wireless
radiotelephones within 1 meter of the AED – turn power OFF to the radiotelephone and other like equipment
near the incident.