5-2
Confidential
8501-00-1800, Revision 05
Candela Corporation
AlexTriVantage Laser System
Introduction
This chapter provides troubleshooting and diagnostic information for the AlexTriVantage
Laser System.
Troubleshooting
These troubleshooting procedures do not replace the instructions or procedures provided in
this guide. Review all instructions and procedures in this guide before performing the
following troubleshooting procedures.
provides troubleshooting information for the AlexTriVantage Laser system.
Table 5-1: Troubleshooting the AlexTriVantage Laser System
Situation/Symptom
Probable Cause or Fault Message
Solution
The system cannot be
turned on properly.
The power is not connected properly.
Check the power cable and the facility circuit
breaker.
The laser system circuit breaker is in the "off"
position.
Switch the circuit breaker to the "on" position.
The keylock switch was not fully engaged.
Turn the keylock switch fully clockwise to the
"S" position and release.
The system will not go into
the Ready state.
WARNING--SECURE REMOTE INTERLOCk
CONNECTOR.
Check the remote interlock connection. Press
any control panel switch to clear the warning
message.
WARNING --RELEASE TRIGGER SWITCH.
The trigger switch must not be pressed when
entering the Ready state. This is a safety
precaution to prevent unexpected pulsing.
Press any control panel switch to clear the
warning message.
PLEASE WAIT--SYSTEM DELAY REQUIRED.
The system has been cycled between Ready
and Standby states too many times over a
short period. Delay prevents HV dump
resistor from overheating. When message is
cleared, system can be put in Ready state. (1
minute wait or less.)
The rate is set at 5 Hz and
the laser pulses at 4 Hz.
The system has degraded.
A preventative maintenance (PM) visit is
required. Call Service.
Coolant over-temperature
WARNING--COOLANT OVER-
TEMPERATURE.
Insure that the system's air circulation is not
blocked. There should be a 15" clearance
between the rear panel and a wall. Remove
any items draped over system. Allow the
system to cool for 10 minutes in the Standby
state before continuing. If problem persists,
turn off system and call Service.
Summary of Contents for AlexTriVantage
Page 14: ...xiv Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 30: ...1 16 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 64: ...2 34 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 82: ...4 8 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 110: ...B 4 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 112: ...C 2 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...