8501-00-1800, Revision 05
Confidential
8-3
Candela Corporation
AlexTriVantage Laser System
Warning: Make sure all personnel in the area are wearing safety eyewear
appropriate for the AlexTriVantage Laser System.
Improper internal calibration of this laser system will cause delivery of lower
or higher fluences and potential burning of patients. This procedure must be
followed precisely for proper results. If the "Final verification of User
Calibration Energy" section fails, contact Candela Technical Support for
further information.
Once the Laser Energy Circuit CAL procedure has been started, the
previously saved CAL parameters will be erased and the Laser Energy Circuit
CAL procedures must be completed in order to use the laser for treatment
again.