Electromagnetic Interference
For information on Electromagnetic Compatibility specifications, see
on page D-1
WARNING!
The AlexTriVantage Laser System should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is necessary, the equipment
should be observed to verify normal operation in the configuration in which it will be
used.
Precautions
General Precautions
To reduce the risk of shock, do not remove covers. Refer servicing to qualified service
personnel.
In the United States, the facility operating a Candela laser system should follow OSHA
guidelines and applicable ANSI standards for the safe use of lasers.
In Canada, Candela laser systems should be installed and operated in accordance with
CAN/CSA -Z386-92: Laser Safety in Health Care Facilities.
In Australia and New Zealand, the facility operating a Candela laser system should be
aware of the requirements of AS/NZS 2211.1, "Laser Safety, Part 1: Equipment
Classification and User's Guide", designed to protect persons from laser radiation
affecting the eye status of users by implementing a thorough system of examination
and reporting.
Laser Room Precautions
Identify the laser room clearly. Post appropriate warning signs in prominent locations at
all entrances to the laser room.
Cover all windows, portholes, etc. with opaque material to prevent unintended viewing
or laser light escaping from the laser room.
When the AlexTriVantage Laser System is in operation, restrict entry and limit access
to the laser room only to personnel that are both essential to the procedure and well
trained in laser safety precautions.
Make sure that all laser room personnel are familiar with the laser system controls and
know how to shut down the laser system instantly in an emergency.
Caution!
The use of flammable anesthetics or oxidizing gases such as nitrous oxide
and oxygen should be avoided. The high temperature produced during normal use of
the laser equipment may ignite some materials, for example cotton or gauze pads
when saturated with oxygen. The solvents of adhesives and flammable solutions used
for cleaning and disinfecting should be allowed to evaporate before the laser
equipment is used. Attention should also be drawn to the danger of ignition of
endogenous gases.
Summary of Contents for AlexTriVantage
Page 14: ...xiv Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 30: ...1 16 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 64: ...2 34 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 82: ...4 8 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 110: ...B 4 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Page 112: ...C 2 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...