Table 7-5: Wiring for RS-232 communications
CH201 connection
CH201
terminal function
Data logger connection
A200 connection
TX
Transmit
C (even) (RX terminal
of a COM port pair)
RX
RX
Receive
C (odd) (TX terminal
of a COM port pair)
TX
COM G
Ground
G
G
DC In 1 or DC In 2
12 VDC
No connection
+12Vdc
FIGURE 7-4. CH201 connected to a CR1000X via RS-232
RS-232 settings are as follows:
Bits per second: 9600
Data bits:
8
Parity:
None
Stop bits:
1
Flow control:
None
CH201 12V Charging Regulator