9.3 Sensor card replacement procedure
CAUTION:
The BaroVUE 10 is sensitive to static when the main case is opened. To avoid damage, take
adequate anti-static measures when handling.
1. Contact Campbell Scientific for your replacement sensor card and take this newly
calibrated card with you to the site of the BaroVUE 10 installation.
CAUTION:
Before handling the BaroVUE 10, wash hands and wear gloves to avoid getting dirt or
grease on the element.
2. Open the main case by unclipping the latch and opening the main door.
FIGURE 9-1. BaroVUE 10 with main door open
BaroVUE™10 Digital Barometer
30