background image

5

Recommendations for use

Before taking any readings, we recommend checking device

settings, especially those regarding the type of medium used.

Remember that the sensor does not have to be connected to

enter all settings. If a change is made during measurement,

wait a few seconds for the device to readjust.
If the device is switched on after connecting the sensor, it will

take less time to display the current flow rate.
To assure the accuracy of the measurement, remove all air

from the system and use suitable strainers to prevent sludge

from building up. 

Changes in the flow rate displayed may be caused by air in

the system.
Handle valves with care and be careful when adjusting and

handling  them  to  prevent  any  excessive  leakages  of  liquid

from the system.
The flow rate meter and sensor are electronic devices which

can  be  damaged  when  dropped,  banged  or  used

inappropriately.

A form is provided at the end of this document, which can be

photocopied and used for future checks.

How to take a reading

1. Shut-off  the  valve  that  the  sensor  is  to  be  inserted  in,

making sure that all other valves are open.

2. Remove the lock, the ring and the cap (only remove the

cap if the valve is shut properly).

3. Insert the sensor holder cap, the ring and lock, checking

the position of each; if the valve is mounted horizontally,

use the discharge screw on the back to help insert the

sensor holder.

4. Fully open the valve with the sensor, using the knob then 

connect the cable to the sensor and to the device.

5. Switch the device on and go to measurement mode.

Wait a few seconds for the flow rate in the circuit to settle and

for any air in the system to be ejected. 

Slowly  regulate  the  opposite  shut-off  valve  and  wait  for

the system to settle at the required value (if the balancing

is  repeated  more  than  once  due  to  their  being  several

branches  in  the  circuit,  extreme  accuracy  is  not

necessary, especially the first few times). 

Once  the  required  flow  rate  has  been  reached,

disconnect the cable from the sensor.

6. Shut-off the sensor holder cap valve using the handle.

7. Remove the lock ring and sensor holder cap.

8. Insert the blind cap into the ball valve and secure the seal

ring with the lock. 

9. Re-open the valve fully using the knob.

The  ball  valve  used  for  regulation  purposes  does  not

behave in a linear way and, for low flow rates in particular,

regulation may be difficult because small changes in the

ball  valve  will  be  associated  with  large  changes  in  the

flow rate. Repeat this for all other manifold branches.

1

6

2

3

4

5

7

8

9

Summary of Contents for GE 130 Series

Page 1: ...luido circolante Lo strumento di misura possiede le informazioni relative alla superficie di passaggio del fluido e i fattori di conversione in tale modo possibile ricavare il valore della portata ist...

Page 2: ...doneo all utilizzo con sostanze potenzialmente pericolose per l uomo come acidi liquidi fortemente basici liquidi infiammabili tossici o esplosivi Utilizzare esclusivamente sensori Caleffi approvati N...

Page 3: ...modo corretto e la portata indicata 0 Il Display LCD permette di visualizzare la portata durante la misura i men di configurazione e le altre informazioni necessarie per il corretto utilizzo dello str...

Page 4: ...ento divenga corretta MENU UNITA UNIT MENU LIQUIDO LIQUID MENU VERSION MODALIT DI MISURA UNITA UNIT l h UNITA UNIT l min UNITA UNIT GPM UNITA UNIT EXIT LIQUIDO LIQUID Acqua Water LIQUIDO LIQUID G30 LI...

Page 5: ...o se la valvola risulta correttamente chiusa 3 Inserire il tappo porta sensore l anello di blocco e il fermo verificandone il corretto posizionamento in caso di montaggio orizzontale della valvola uti...

Page 6: ...ianto Tipo Potenza Dati Portata impianto Temperature estive Temperature invernali Fluido Portata circuito 1 Portata circuito 2 Portata circuito 3 Dati Portata circuito 4 collettore Portata circuito 5...

Page 7: ...measured The flow meter contains data regarding the flowing surface of the medium and conversion factors needed to be able to calculate the instantaneous flow rate The device has an easy grip and non...

Page 8: ...d be hazardous to humans such as acids highly basic liquids and toxic flammable or explosive liquids Only use Caleffi approved sensors Do not open the flow meter or sensor if tampered with the warrant...

Page 9: ...than 300 l h if the sensor is not connected or inserted properly the device will not be able to measure correctly and a flow rate of 0 will be shown The LCD display shows flow rate during measurement...

Page 10: ...seconds for the reading to be corrected MENU UNIT MENU LIQUID MENU VERSION MEASUREMENT MODE UNIT l h UNIT l min UNIT GPM UNIT EXIT LIQUID Water LIQUID G30 LIQUID G50 LIQUID EXIT CALEFFI S p A I 001 CA...

Page 11: ...cap if the valve is shut properly 3 Insert the sensor holder cap the ring and lock checking the position of each if the valve is mounted horizontally use the discharge screw on the back to help inser...

Page 12: ...tem Type Power System Flow rate data Summer temperatures Winter temperatures Medium Circuit 1 flow rate Circuit 2 flow rate Circuit 3 flow rate Manifold Circuit 4 flow rate data Circuit 5 flow rate Ci...

Page 13: ...des Mediums bestimmt werden kann Das Messger t verf gt ber die Informationen zur Str mungsfl che des Messmediums und den Konversionsfaktoren Dadurch l sst sich der momentane Durchflusswert ermitteln D...

Page 14: ...bzw andere Fl ssigkeiten tauchen Die Sonde ist nicht f r den Einsatz mit potenziell f r den Menschen gef hrlichen Stoffen wie S uren stark basischen Fl ssigkeiten entz ndlichen gesundheitssch dlichen...

Page 15: ...ne vorschriftsm ige Messung ausf hren und am Display wird der Durchfluss 0 angezeigt Am LCD Display werden die Durchflussmenge beim Messvorgang die Konfigurationsmen s und die anderen Informationen f...

Page 16: ...einige Sekunden auf die Aktualisierung der Ger teanzeige warten MENU UNITA UNIT MENU LIQUIDO LIQUID MENU VERSION MESS MODUS UNITA UNIT l h UNITA UNIT l min UNITA UNIT GPM UNITA UNIT EXIT LIQUIDO LIQU...

Page 17: ...ndenverschluss den Ring und die Sicherung vorschriftsm ig einsetzen bei horizontaler Installation des Ventils die entsprechende Schraube auf der R ckseite als Einbauhilfe des Sondenverschlusses benutz...

Page 18: ...Typ Leistung Anlagen Durchfluss Daten Sommertemperaturen Wintertemperaturen Fl ssigmedium Durchfluss Kreis 1 Durchfluss Kreis 2 Durchfluss Kreis 3 Verteiler Durchfluss Kreis 4 Daten Durchfluss Kreis 5...

Reviews: