Service Manual
24 | MN260 B
7.3 REPROGRAMMING AFTER OVERHAUL AND REPAIR PROCEDURES
A flow calibration procedure is required after replacing any of the following components or assemblies: ATF Module, Com-
pressor and Product Tank, PCB Circuit Board, Proportional Valve. If any of these components are replaced, flow calibration
is required. Flow calibration can be done at a CAIRE authorized service center or in the field using item T-10560. See manual
MN053 for details on performing flow calibration and firmware updates in the field using item T-10560. Equipment or procedures
to perform a flow calibration are not contained in this manual.
NOTE: Further damage to the SAROS or other components is possible if the flow calibration is not performed.
7.4 SAROS MAIN COMPONENTS
PCB Assembly
Part #SP9710V2-SEQ
Inlet Cap Assembly
Part # SP9756V2-SEQ
Compressor Assembly
Part #SP9238-SEQ
Battery Bridge Port Wire Harness
Assembly
Part # 21494875
ATF
Assembly
Part # SP9729-SEQ
Figure 7-1. Inside SAROS