PN 21033264 B | User Manual    5 - ENG

HELiOS 300 & 800

Introduction

A liquid oxygen system is designed to provide sup-

plementary oxygen as prescribed by a physician. 

Your liquid oxygen system includes a HELiOS 

portable unit and a stationary reservoir.    The res-

ervoir is the larger container that is used to fill your 

HELiOS portable. This user manual contains only 

the instructions for using the HELiOS portable. Re-

fer to the user manual supplied with your reservoir 

unit for information on its operation.

HELiOS portable liquid oxygen units are designed 

to be an ambulatory source of oxygen for an 

extended period of time. They are filled with liquid 

oxygen from a reservoir unit with a compatible fill 

connector. HELiOS portables are available in two 

different sizes.  

The H300 Plus is primarily a conserving device 

and holds 0.38 Liters of liquid oxygen. It provides 

demand (pulse) flow settings 1-4, and limited low 

dose continuous flow settings at 0.75 LPM and 

below only. 

The H850 Marathon can be used as a conserving 

or continuous flow device and holds 0.84 liters of 

liquid oxygen. It provides continuous flow from 1-6 

LPM and demand flow on settings of 1-4.

H300 and H850 portable  

units shown.

Specifications

•  Mode of Operation:  Demand and Continu-

ous Flow

•  Type of Protection Against Electrical Shock: 

Internally Powered Equipment

• 

 Degree of Protection Against Electri-

cal Shock: Type BF Applied Part

•  IPX1 Classification According to the Degree 

of Protection Against Ingress of Water: Drip 

Proof

• 

 Equipment not suitable for use in the 

presence of flammable mixtures

HELiOS Plus  
(H300) 

HELiOS 
Marathon 
(H850)

LOX capacity 

0.9 lb (0,41 
kg) 

2.0 lb (0,91 kg)

Gaseous 
equivalent 
capacity 

308 L 

685 L

Weight, filled 

3.6 lb (1,63 
kg) 

5.6 lb (2,54 kg)

Weight, empty  2.7 lb (1,22 

kg) 

3.6 lb (1,63 kg)

Height includ-
ing handle 

10.5” (267 
mm) 

15.0” (381 mm)

Typical use 
time at 2 LPM 
demand mode

9 hours 

22 hours

Typical use 
time at 2 LPM 
continuous 
mode

N/A 

6.1 hours

Operating 
pressure 

22 psi (152 
kPa) 

22 psi (152 kPa)

Normal evapo-
ration rate 

1.0 lb/day 
(0,45 kg/day) 

0.85 lb/day 
(0,39 kg/day)

Standard flow 
control range 

Continuous: 
off, .12, .25, .5, 
.75 Demand: 
off, 1, 1.5, 2, 
2.5, 3, 3.5, 4

Continuous: off, 
1,2,3,4,5,6
Demand: off, 
1.5, 2, 2.5, 3, 4

Summary of Contents for HELiOS Marathon

Page 1: ...HELiOSTM Series User Manual English...

Page 2: ...prohibited Reg P003 Do not smoke near unit or while operating unit Reg P002 Type BF applied part degree of protection against electric shock Reg 5333 Warning Reg W001 Council Directive 93 42 EEC conc...

Page 3: ...ctive 2012 19 EU waste electrical and electronic equipment WEEE WEEE IEC 60601 1 Medical electrical equipment Part 1 General requirements for basic safety and essential performance Drip Proof This pro...

Page 4: ...AUTIOUSLY RETURN THE UNIT INTO AN UPRIGHT POSITION IF POSSIBLE IF ANY LIQUID OXYGEN IS ESCAPING LEAVE THE AREA IMME DIATELY AND CALL YOUR HEALTH CARE PROVIDER DO NOT ATTEMPT TO MOVE THE UNIT OR STOP T...

Page 5: ...d as a conserving or continuous flow device and holds 0 84 liters of liquid oxygen It provides continuous flow from 1 6 LPM and demand flow on settings of 1 4 H300 and H850 portable units shown Specif...

Page 6: ...rap 5 Contents Indicator Window 6 Vent Valve Lever 7 Fill Connector QDV Fill Connector Compatibility The H300 and H850 can be filled from any CAIRE top fill TF reservoir that has a CAIRE style male fi...

Page 7: ...through top port Breath is detected through bottom port Operating Instructions Filling 1 Clean the fill connectors on the H300 H850 and the reservoir with a clean dry lint free cloth WARNING THE FILL...

Page 8: ...T LEAVE THE AREA AND CALL YOUR HEALTH CARE PROVIDER IMMEDIATELY WARNING IF LARGE AMOUNTS OF VAPOR ARE ISSUING FROM THE UNITS DURING FILLING STOP FILLING LEAVE THE ROOM AND CALL YOUR HEALTH CARE PROVID...

Page 9: ...continuous flow the H300 H850 will deliver a steady stream of oxygen You must continue to breathe through the nasal cannula normally to continue receiving continuous flow Note Due to the possibility...

Page 10: ...pres ent from humidified exhaled gas Replacing the cannula with a dry one enhances the ability of the Portable to pulse Ensure that the cannula is not kinked Ensure that the cannula tips remain in you...

Page 11: ...r always wipe the male fill connector on the Reservoir and the female fill connector on the Portable with a clean dry cloth before filling NOTE It is recommended that routine maintenance be performed...

Page 12: ...with no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 3010...