background image

PN 21033264 B | User Manual    11 - ENG

HELiOS 300 & 800

Cleaning Standard

WARNING:

 DO NOT ALLOW WA-

TER INTO ANY OF THE CONTROLS, 
THE FILL CONNECTOR, OR THE 
OXYGEN SUPPLY CONNECTOR. 

Caution:

 Clean only after the unit 

is empty and vented.

•  Clean using a solution of mild dish washing 

detergent and water.

•  Apply cleaning solution directly to a lint-free 

cloth. Do not spray cleaners directly on the 

H300/H850.

•  Wipe the outside surface with the lint-free 

cloth until the outside surface is clean.

Caution:

 Do not use high temper-

ature and high pressure washing 
equipment to clean these units.

•  Do not get cleaner on any internal compo-

nents or valves.

•  Allow the unit to dry thoroughly before us-

ing.

adequate pressure for accurate oxygen flow. 

If needed, use an alternate source of oxygen 

such as a flow control valve attached to the 

Reservoir.

6. The Portable does not disengage easily from the 

Reservoir after filling.

• The Portable and Reservoir fill connectors 

may have become frozen. DO NOT USE 

FORCE. Allow a few minutes for the frozen 

parts to warm,  then disengage the Portable 

when the ice has melted. To prevent the units 

from freezing together, always wipe the male 

fill connector on the Reservoir and the female 

fill connector on the Portable with a clean, 

dry cloth before filling.

NOTE:  It is recommended that routine maintenance 
be performed annually on this equipment. Please 
contact your Health Care Provider to ensure this 
routine maintenance is being completed.

WEEE and RoHS

This symbol is to remind the equipment 

owners to return it to a recycling facility 

at the end of its life, per Waste Electrical 

and Electronic Equipment (WEEE) 

Directive.

Our products will comply with the restriction of 

Hazardous Substances (RoHS) directive.  They 

will not contain more than trace amounts of lead or 

other hazardous material content.

Transport and Storage

The device should be stored in the upright position 

well ventilated. Do not allow the device to be tilted 

forward or laying on its front. Operating tem-

perature range of 14°F to 104°F (-10°C to 40°C). 

Storage temperature ranges from -40°F to 158°F 

(-40°C to 70°C).

Maintenance

Clean the fill connectors on both the stationary 

and portable units with a clean, dry, lint-free cloth 

between each fill to prevent freezing and possible 

equipment failure.

NOTE:  Any additional maintenance needed must 
be done by a qualified service technician or service 
provider.

Accessories 

Backpack —HELiOS  H850 

Marathon

Hip pack —HELiOS H300

H850 Cart

Summary of Contents for HELiOS Marathon

Page 1: ...HELiOSTM Series User Manual English...

Page 2: ...prohibited Reg P003 Do not smoke near unit or while operating unit Reg P002 Type BF applied part degree of protection against electric shock Reg 5333 Warning Reg W001 Council Directive 93 42 EEC conc...

Page 3: ...ctive 2012 19 EU waste electrical and electronic equipment WEEE WEEE IEC 60601 1 Medical electrical equipment Part 1 General requirements for basic safety and essential performance Drip Proof This pro...

Page 4: ...AUTIOUSLY RETURN THE UNIT INTO AN UPRIGHT POSITION IF POSSIBLE IF ANY LIQUID OXYGEN IS ESCAPING LEAVE THE AREA IMME DIATELY AND CALL YOUR HEALTH CARE PROVIDER DO NOT ATTEMPT TO MOVE THE UNIT OR STOP T...

Page 5: ...d as a conserving or continuous flow device and holds 0 84 liters of liquid oxygen It provides continuous flow from 1 6 LPM and demand flow on settings of 1 4 H300 and H850 portable units shown Specif...

Page 6: ...rap 5 Contents Indicator Window 6 Vent Valve Lever 7 Fill Connector QDV Fill Connector Compatibility The H300 and H850 can be filled from any CAIRE top fill TF reservoir that has a CAIRE style male fi...

Page 7: ...through top port Breath is detected through bottom port Operating Instructions Filling 1 Clean the fill connectors on the H300 H850 and the reservoir with a clean dry lint free cloth WARNING THE FILL...

Page 8: ...T LEAVE THE AREA AND CALL YOUR HEALTH CARE PROVIDER IMMEDIATELY WARNING IF LARGE AMOUNTS OF VAPOR ARE ISSUING FROM THE UNITS DURING FILLING STOP FILLING LEAVE THE ROOM AND CALL YOUR HEALTH CARE PROVID...

Page 9: ...continuous flow the H300 H850 will deliver a steady stream of oxygen You must continue to breathe through the nasal cannula normally to continue receiving continuous flow Note Due to the possibility...

Page 10: ...pres ent from humidified exhaled gas Replacing the cannula with a dry one enhances the ability of the Portable to pulse Ensure that the cannula is not kinked Ensure that the cannula tips remain in you...

Page 11: ...r always wipe the male fill connector on the Reservoir and the female fill connector on the Portable with a clean dry cloth before filling NOTE It is recommended that routine maintenance be performed...

Page 12: ...with no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 3010...

Reviews: