8 - ENG    PN 21033264 B | User Manual

HELiOS 300 & 800

7. You may see vapor around the fill connection.  

This is normal due to the great difference in 

temperature between the liquid oxygen and the 

warm surrounding air.

8. Maintain a downward force on the H300/H850 

with one hand during the entire filling process 

and keep the vent valve lever open with the other 

hand.

9. Approximately 20-30 seconds into the fill, close 

and re-open the vent valve one or more times.  

This will prevent the valve from freezing open.

10. The unit is full when the hissing sound changes 

in tone and a dense white vapor is seen coming 

from under the shroud of the reservoir.  When 

this is observed, close the vent valve by raising 

the vent valve lever. This will terminate the fill.

11. Hold the handle on the H300/H850 and push 

the release button on the reservoir until the por-

table separates (release button varies by model 

of reservoir).

12. Check the contents indicator to verify that the 

unit is full.

WARNING:

 DO NOT LEAVE UNIT 

UNATTENDED WHILE FILLING. 

WARNING:

 DO NOT OPERATE THE 

H300/H850 WHILE ATTACHED TO 
THE RESERVOIR UNIT.

WARNING:

 SHOULD LEAKAGE BE EXCESSIVE TO 

THE POINT THAT A STREAM OF LIQUID IS PRES-
ENT, LEAVE THE AREA AND CALL YOUR HEALTH 
CARE PROVIDER IMMEDIATELY.

WARNING:

 IF LARGE AMOUNTS OF VAPOR ARE 

ISSUING FROM THE UNITS DURING FILLING, 
STOP FILLING, LEAVE THE ROOM AND CALL 
YOUR HEALTH CARE PROVIDER.

WARNING: 

IF PROLONGED HISSING IS HEARD, 

STOP USE AND CONTACT YOUR HEALTH CARE 
PROVIDER IMMEDIATELY

WARNING:

 IF THE H300/H850 DOES NOT SEPA-

RATE EASILY, DO NOT USE FORCE.  THE UNITS 
MAY BE FROZEN TOGETHER.  LEAVE THE UNITS 
CONNECTED AND WAIT UNTIL THEY WARM UP 
– THEN THEY WILL SEPARATE EASILY.  DO NOT 
TOUCH ANY FROSTED PARTS.

Caution:

 Should there be any 

liquid leakage from the portable 
after separating the units, , set 
the H300/H850 aside, ensuring it 
remains vertical, leave the room, 
and call your health care provider 
immediately.

Caution:

 Should there be any liquid leakage 

from the reservoir after separating the units, 
open windows in the room, leave the room im-
mediately, and call your health care provider.

Caution:

 Check the liquid level only after the 

vent valve is closed.

Caution:

 If the H300/H850 is being refilled 

immediately after a period of use, fill it, and 
then wait approximately one (1) hour after fill-
ing with the flow control knob set to off (“0”) 
for the pressure to stabilize.  Then operate as 
normal.

NOTE:  If the HELiOS portable is warm or has not 
been filled recently, it may take approximately 90 
seconds to fill an H300 or approximately four (4) 
minutes to fill an H850.

Summary of Contents for HELiOS Marathon

Page 1: ...HELiOSTM Series User Manual English...

Page 2: ...prohibited Reg P003 Do not smoke near unit or while operating unit Reg P002 Type BF applied part degree of protection against electric shock Reg 5333 Warning Reg W001 Council Directive 93 42 EEC conc...

Page 3: ...ctive 2012 19 EU waste electrical and electronic equipment WEEE WEEE IEC 60601 1 Medical electrical equipment Part 1 General requirements for basic safety and essential performance Drip Proof This pro...

Page 4: ...AUTIOUSLY RETURN THE UNIT INTO AN UPRIGHT POSITION IF POSSIBLE IF ANY LIQUID OXYGEN IS ESCAPING LEAVE THE AREA IMME DIATELY AND CALL YOUR HEALTH CARE PROVIDER DO NOT ATTEMPT TO MOVE THE UNIT OR STOP T...

Page 5: ...d as a conserving or continuous flow device and holds 0 84 liters of liquid oxygen It provides continuous flow from 1 6 LPM and demand flow on settings of 1 4 H300 and H850 portable units shown Specif...

Page 6: ...rap 5 Contents Indicator Window 6 Vent Valve Lever 7 Fill Connector QDV Fill Connector Compatibility The H300 and H850 can be filled from any CAIRE top fill TF reservoir that has a CAIRE style male fi...

Page 7: ...through top port Breath is detected through bottom port Operating Instructions Filling 1 Clean the fill connectors on the H300 H850 and the reservoir with a clean dry lint free cloth WARNING THE FILL...

Page 8: ...T LEAVE THE AREA AND CALL YOUR HEALTH CARE PROVIDER IMMEDIATELY WARNING IF LARGE AMOUNTS OF VAPOR ARE ISSUING FROM THE UNITS DURING FILLING STOP FILLING LEAVE THE ROOM AND CALL YOUR HEALTH CARE PROVID...

Page 9: ...continuous flow the H300 H850 will deliver a steady stream of oxygen You must continue to breathe through the nasal cannula normally to continue receiving continuous flow Note Due to the possibility...

Page 10: ...pres ent from humidified exhaled gas Replacing the cannula with a dry one enhances the ability of the Portable to pulse Ensure that the cannula is not kinked Ensure that the cannula tips remain in you...

Page 11: ...r always wipe the male fill connector on the Reservoir and the female fill connector on the Portable with a clean dry cloth before filling NOTE It is recommended that routine maintenance be performed...

Page 12: ...with no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 3010...

Reviews: