Oxygen Concentrator 

Technical Service Manual 

| PN 20631679 Rev G       

7

 

 

 

 

ECLIPSE 5 OXYGEN SYSTEM SPECIFICATIONS

 

Oxygen Concentrator

 

 

Dimensions (H x W X D)

 

19.3 x 12.3 x 7.1 inches

 

(49.0cm x 31.2cm x 18.0cm)

 

Weight

 

Eclipse 5

 

Power Cartridge

 

 

15.0 pounds

 

3.4 pounds

 

Flow Settings

 

Continuous Flow (measured in Liters Per Minute LPM)

 

Pulse 

Dose (measured in mL)

 

 

0.5 to 3.0 LPM (0.5 liter increments)

 

 

Settings 1.0-6.0: 16mL, 24mL, 32mL, 40mL, 48mL, 56mL, 64mL, 72mL, 80mL, 88mL, 96mL 
Settings 7-9: 128mL, 160mL, 192mL

 

Continuous Flow Accuracy

 

+/- 10% or 200ml/min, whichever is greater

 

Oxygen Concentration

 

87

95.5% for all flow settings

 

Maximum System Pressure

 

14 psig (96.5 kPa)

 

Oxygen Output Pressure

 

5.0 psig (34.5 kPa) nominal

 

Oxygen Concentration

 

Status Indicator

 

Green Light = Normal Operation

 

Yellow Light = Warning or Caution, less than 85% ± 3%

 

Red Light Flashing = Abnormal Operation, less than 70% ± 5% (Note: only for 3-LED models)

 

Nominal Sound Level

 

3.0 LPM Continuous Flow

 

3.0 Pulse Dose Mode

 

 

48 dB(A)

 

40 dB(A)

 

Less than 57 dB while alarming

 

Operating Environment 

Temperature

 

Humidity

 

 

50º F to 104º F (+10° to 40°C)

 

10% to 95%, Non-condensing, 82.4°F (28°C) Maximum Dew point

 

Storage Environment Temperature

 

Humidity

 

 

-4º F to 140º F (-20° to 60°C)

 

Up to 95% Non-condensing

 

Altitude

 

 13,123 feet (0 - 4,000 meters)

 

Nominal Power

 

0.5 LPM Continuous flow 3 LPM 
Continuous Flow

 

1.0 Pulse Dose Setting

 

6.0 Pulse Dose Setting Battery 
Charging

 

 

48 Watts

 

145 Watts

 

45 Watts

 

95 Watts

 

50 Watts

 

Nominal Power Cartridge operating time* Continuous 

Flow

 

2.0 LPM

 

3.0 LPM

 

Pulse Dose (@ 12 BPM)

 

Setting 2.0

 

Setting 6.0

 

 

2.0 hours

 

1.3 hours

 

 

5.2 hours

 

3.5 hours

 

* Note: Times will decrease with higher bolus size, breath rate, ambient temperature, power 
cartridge age and use over time.

 

Outlet Pressure

 

Nominal: 7.0 psig

 

Maximum: 14.0 psig

 

Continuous Flow Indication

 

Expressed in liters per minute (LPM)

 

Audible Alarm Indicators

 

 

Loss of Power/Hot Power Cartridge

 

 

Low Power Cartridge/Warm Power Cartridge

 

 

Low Oxygen Output

 

 

O

Flow Outside Normal Limits

 

 

Unit Malfunction

 

 

Low 9V battery (Note: only for 3-LED model)

 

Summary of Contents for Eclipse 5 with autoSAT

Page 1: ...Eclipse 5 with autoSAT Oxygen Concentrator TECHNICAL SERVICE MANUAL...

Page 2: ...AC Power Supply 18 DC Power Supply 18 Power Cartridge 19 Typical New Power Cartridge Operation Time 20 Charging Algorithm 22 Training The Patient 23 Introduction 23 Pre Delivery Check List 23 Indicati...

Page 3: ...ication Tests 48 Record Hours of Operation Software Version 49 Electrical Safety Test 50 Cleaning the Eclipse 5 50 Provider Service and Maintenance Record 51 Shipping and Transporting the Eclipse 5 52...

Page 4: ...rinciples of operation Providers are given proper guidance in the use of the Eclipse 5 and its accessories that can beconveyed to patients Providers are made aware of the care diagnostics maintenance...

Page 5: ...rd commercial aircraft Flow Setting Indicator Pulse Mode Operation Device operating normally power button Increase Flow Setting Decrease Flow Setting ISO 7000 Graphical symbols for use on equipment In...

Page 6: ...RIDGE BATTERY PN 7082 SEQ PN 20852326S with IEC type A cord 20852325S with IEC type C cord and 20852324S with IEC type G cord two LED model Oxygen Outlet Port Control Panel FAA Approval Icon Rating La...

Page 7: ...0 LPM Continuous Flow 3 0 Pulse Dose Mode 48 dB A 40 dB A Less than 57 dB while alarming Operating Environment Temperature Humidity 50 F to 104 F 10 to 40 C 10 to 95 Non condensing 82 4 F 28 C Maximu...

Page 8: ...w at the last Continuous Flow setting After another 15 seconds the Eclipse 5 stops delivering Continuous Flow and waits 15 seconds for a breath The Eclipse 5 will stay in this modality until a breath...

Page 9: ...gnetic Compatibility Emissions EN55011 FCC Part J15 Class B Conducted EN55011 FCC Part J15 Class B Radiated EN61000 3 2 EN61000 3 3 EN61000 4 2 EN61000 4 3 EN61000 4 4 EN61000 4 5 EN61000 4 6 EN61000...

Page 10: ...provided in this manual Portable and mobile RF communications equipment can affect Medical Electrical Equipment The use of Accessories transducers and cables other than those specified with the excep...

Page 11: ...ons it is recommended that the Eclipse 5 is powered from an uninterruptible power supply UPS Power frequency magnetic field IEC 61000 4 8 30A m Power frequency magnetic fields should be at levels char...

Page 12: ...nd controls both the power source and the Power Cartridge charger Figure 1 Eclipse 5 System Block Diagram ATF CONCENTRATOR MODULE The Eclipse 5 uses a passive system to separate oxygen from air Air fl...

Page 13: ...sor enclosure serves to efficiently draw cooling air in over the compressor cylinders while simultaneously pushing exhaust gas out of the concentrator POWER DISTRIBUTION The Power Manager takes extern...

Page 14: ...ng has been saved and will be used at start up Power Cartridge battery Status Gauge This indicator displays the charge remaining in the battery Each of the five horizontal gray bars represents approxi...

Page 15: ...y rates Regardless of setting the pulse profile is a simple square wave pulse based on a 16 mL volume Refer to Pulse Profiles table below Pulse durations are no less than 100 milli seconds at the begi...

Page 16: ...S x PS 1 is the most sensitive setting PS 3 is the least sensitive setting Factory default is PS 2 HOURS OF OPERATION The Eclipse 5 hour meter provides valuable information on hours of operation Provi...

Page 17: ...t needs EDAT software is capable of sensing and displaying historical event logs and values of components and sensors within the Eclipse 5 system This can be used to determine system faults or user er...

Page 18: ...t is illuminated when the AC Power Supply is supplying 28VDC power If the LED is not illuminated there is no input power available In addition the Power Supply contains protection circuits for output...

Page 19: ...f the user If the Eclipse 5 power cartridge is being charged the battery symbol will display with a waterfall animation However if the power cartridge is not charging then the battery symbol will rema...

Page 20: ...d with the Eclipse 5 is not fully charged Before using the Eclipse 5 Oxygen System for the first time the Power Cartridge must be fully charged Store the power cartridge in a cool dry location Do not...

Reviews: