background image

Oxygen Concentrator 

Technical Service Manual 

| PN 20631679 Rev G       

20

 

 

 

 

The Eclipse 5 Power Cartridge operation time may be affected by several factors such as bolus size, breathing rates, ambient 
temperature, age of power cartridge and use over time. The table below describes the typical operating time for a new Power 
Cartridge.

 

 

If the Eclipse 5 is used in Pulse Dose Mode, there will be longer operating time. The patient should consult their physician 
for a prescription for Pulse Dose Mode.

 

 

TYPICAL NEW POWER CARTRIDGE OPERATING 
TIMES

 

(At  nominal 

temperature 

of 25ºC/77°F)

 

 

0.5 LPM 

4.4 hours 

-

 

1.0 LPM 

3.7 hours 

16mL 1.0 

5.4 hours

 

 

2.0 LPM 

2.0 hours 

32mL 2.0 

5.1 hours

 

3.0 LPM 

1.3 hours 

48mL 3.0 

4.9 hours

 

 

64mL 4.0 

4.0 hours

 

80mL 5.0 

3.7 hours

 

 

96mL 6.0 

3.5 hours

 

  128mL 7 

2.5 hours

 

 

 160mL 8 

2.0 hours

 

 

 192mL 9 

1.7 hours

 

 

 

Battery Consumption Chart

 

NOTE

: Battery times will decrease with higher bolus size, breath rate, ambient temperature, Battery age and use over time.

 

 

The Power Cartridge packaged with the Eclipse 5 is not fully charged. Before using the Eclipse 5 Oxygen System for the 
first time, the Power Cartridge must be fully charged.

 

 

Store the power cartridge in a cool, dry location. Do not leave the Eclipse 5 or Power Cartridge in a vehicle or trunk during 
a hot or cold day.

 

 

When checking the Eclipse 5 as baggage on a commercial airline flight for international travel, remove the Power Cartridge 
and properly package.

 

When shipping the Eclipse 5 for any reason, remove the Power Cartridge from the Eclipse 5.

 

 

If the Power Cartridge gets too warm, charging will not begin until the Power Cartridge sufficiently cools. Consider 
removing the Power Cartridge to allow for faster cooling.

 

 

The Power Cartridge operating time is longer if the Eclipse 5 is operated in Pulse Dose Mode. (Refer to Battery 
Consumption Chart above.)

 

 

The typical time to recharge the Power Cartridge to achieve 80% capacity from a fully discharged Power Cartridge is 1.8 
hours to 5.0 hours, dependent upon the flow setting.

 

The capacity of the Eclipse 5 Power Cartridge is determined by electronics and the Eclipse 5 software.

 

Battery Duration

 

(12 BPM)

 

Pulse 

Setting

 

Battery Duration

 

Continuous 

Flow Setting

 

Summary of Contents for Eclipse 5 with autoSAT

Page 1: ...Eclipse 5 with autoSAT Oxygen Concentrator TECHNICAL SERVICE MANUAL...

Page 2: ...AC Power Supply 18 DC Power Supply 18 Power Cartridge 19 Typical New Power Cartridge Operation Time 20 Charging Algorithm 22 Training The Patient 23 Introduction 23 Pre Delivery Check List 23 Indicati...

Page 3: ...ication Tests 48 Record Hours of Operation Software Version 49 Electrical Safety Test 50 Cleaning the Eclipse 5 50 Provider Service and Maintenance Record 51 Shipping and Transporting the Eclipse 5 52...

Page 4: ...rinciples of operation Providers are given proper guidance in the use of the Eclipse 5 and its accessories that can beconveyed to patients Providers are made aware of the care diagnostics maintenance...

Page 5: ...rd commercial aircraft Flow Setting Indicator Pulse Mode Operation Device operating normally power button Increase Flow Setting Decrease Flow Setting ISO 7000 Graphical symbols for use on equipment In...

Page 6: ...RIDGE BATTERY PN 7082 SEQ PN 20852326S with IEC type A cord 20852325S with IEC type C cord and 20852324S with IEC type G cord two LED model Oxygen Outlet Port Control Panel FAA Approval Icon Rating La...

Page 7: ...0 LPM Continuous Flow 3 0 Pulse Dose Mode 48 dB A 40 dB A Less than 57 dB while alarming Operating Environment Temperature Humidity 50 F to 104 F 10 to 40 C 10 to 95 Non condensing 82 4 F 28 C Maximu...

Page 8: ...w at the last Continuous Flow setting After another 15 seconds the Eclipse 5 stops delivering Continuous Flow and waits 15 seconds for a breath The Eclipse 5 will stay in this modality until a breath...

Page 9: ...gnetic Compatibility Emissions EN55011 FCC Part J15 Class B Conducted EN55011 FCC Part J15 Class B Radiated EN61000 3 2 EN61000 3 3 EN61000 4 2 EN61000 4 3 EN61000 4 4 EN61000 4 5 EN61000 4 6 EN61000...

Page 10: ...provided in this manual Portable and mobile RF communications equipment can affect Medical Electrical Equipment The use of Accessories transducers and cables other than those specified with the excep...

Page 11: ...ons it is recommended that the Eclipse 5 is powered from an uninterruptible power supply UPS Power frequency magnetic field IEC 61000 4 8 30A m Power frequency magnetic fields should be at levels char...

Page 12: ...nd controls both the power source and the Power Cartridge charger Figure 1 Eclipse 5 System Block Diagram ATF CONCENTRATOR MODULE The Eclipse 5 uses a passive system to separate oxygen from air Air fl...

Page 13: ...sor enclosure serves to efficiently draw cooling air in over the compressor cylinders while simultaneously pushing exhaust gas out of the concentrator POWER DISTRIBUTION The Power Manager takes extern...

Page 14: ...ng has been saved and will be used at start up Power Cartridge battery Status Gauge This indicator displays the charge remaining in the battery Each of the five horizontal gray bars represents approxi...

Page 15: ...y rates Regardless of setting the pulse profile is a simple square wave pulse based on a 16 mL volume Refer to Pulse Profiles table below Pulse durations are no less than 100 milli seconds at the begi...

Page 16: ...S x PS 1 is the most sensitive setting PS 3 is the least sensitive setting Factory default is PS 2 HOURS OF OPERATION The Eclipse 5 hour meter provides valuable information on hours of operation Provi...

Page 17: ...t needs EDAT software is capable of sensing and displaying historical event logs and values of components and sensors within the Eclipse 5 system This can be used to determine system faults or user er...

Page 18: ...t is illuminated when the AC Power Supply is supplying 28VDC power If the LED is not illuminated there is no input power available In addition the Power Supply contains protection circuits for output...

Page 19: ...f the user If the Eclipse 5 power cartridge is being charged the battery symbol will display with a waterfall animation However if the power cartridge is not charging then the battery symbol will rema...

Page 20: ...d with the Eclipse 5 is not fully charged Before using the Eclipse 5 Oxygen System for the first time the Power Cartridge must be fully charged Store the power cartridge in a cool dry location Do not...

Reviews: