background image

Oxygen Concentrator 

Technical Service Manual 

| PN 20631679 Rev G       

18

 

 

 

 

POWER SUPPLIES

 

The  Eclipse  5  may  operate  from  either  the  AC  or  DC  Power  Supply  or  the  Power  Cartridge.  When  power  is 
available from an external supply, the Eclipse 5 will draw from the external source rather than depleting the Power 
Cartridge. Connection to external power is indicated when the External Power Present Indicator located on the 
Control Panel is illuminated.

 

 
 

AC POWER SUPPLY

 

The AC Power Supply is a universal input type, capable of accepting 

100-240VAC, 50/60 Hz

. It is capable of delivering up to 200W of 

28VDC output. The input cord requires a grounded receptacle. Country specific cords or universal power adapter PN 5591-SEQ may be 

used with the AC Power Supply, as the power supply is equipped with a universal input receptacle. When used in a medical care facility, 

international safety standards require the use of hospital grade AC power cord with the Eclipse 5.

 

 

 

 

Figure 5: Eclipse 5 AC Power Supply

 

 

The Eclipse 5 AC Power Supply does not contain a fan. When in use, the AC Power Supply should be located in a well-ventilated area.

 

 

Located near the output cord, there is a green LED that is illuminated when the AC Power Supply is supplying 28VDC power. If the LED is 

not illuminated, there is no input power available. In addition, the Power Supply contains protection circuits for output over-current, input 

over-voltage, and internal over-temperature conditions. If any of these conditions exist, output power will be interrupted and the LED turns 

off. However, these three conditions are self-resetting, such that output power will resume when protection circuits fall back into acceptable 

operating ranges.

 

Note:

 

Eclipse 5 models with 3 LEDs can use the newer 2-pin power supply (PN: 20852326S (with IEC type A cord), 20852325S (with IEC type 

C cord), and 20852324S (with IEC type G cord)). In order for the Eclipse 5 models with 2 LEDAs to be used with the older 3-pin power 

supply (PN 5941-SEQ), the CSA labeling would need be removed from the unit. The applied part would then change from type BF to type 

B.

 

 
 

DC POWER SUPPLY

 

The DC Power Supply is intended for use with DC accessory outlets, such as those found in motor vehicles. Input voltage to the DC 

Power Supply is 11.5-16 VDC. Sized to be compatible with most passenger vehicle electrical systems, it is capable of delivering up to 

150W of 26VDC output to the Eclipse 5. 

 
 
 
 

 

 

PN: 5941-SEQ (three LED model)

 

PN: 20852326S (with IEC type A cord), 20852325S (with 

IEC type C cord), and 20852324S (with IEC type G cord) 

(two LED model) 

Summary of Contents for Eclipse 5 with autoSAT

Page 1: ...Eclipse 5 with autoSAT Oxygen Concentrator TECHNICAL SERVICE MANUAL...

Page 2: ...AC Power Supply 18 DC Power Supply 18 Power Cartridge 19 Typical New Power Cartridge Operation Time 20 Charging Algorithm 22 Training The Patient 23 Introduction 23 Pre Delivery Check List 23 Indicati...

Page 3: ...ication Tests 48 Record Hours of Operation Software Version 49 Electrical Safety Test 50 Cleaning the Eclipse 5 50 Provider Service and Maintenance Record 51 Shipping and Transporting the Eclipse 5 52...

Page 4: ...rinciples of operation Providers are given proper guidance in the use of the Eclipse 5 and its accessories that can beconveyed to patients Providers are made aware of the care diagnostics maintenance...

Page 5: ...rd commercial aircraft Flow Setting Indicator Pulse Mode Operation Device operating normally power button Increase Flow Setting Decrease Flow Setting ISO 7000 Graphical symbols for use on equipment In...

Page 6: ...RIDGE BATTERY PN 7082 SEQ PN 20852326S with IEC type A cord 20852325S with IEC type C cord and 20852324S with IEC type G cord two LED model Oxygen Outlet Port Control Panel FAA Approval Icon Rating La...

Page 7: ...0 LPM Continuous Flow 3 0 Pulse Dose Mode 48 dB A 40 dB A Less than 57 dB while alarming Operating Environment Temperature Humidity 50 F to 104 F 10 to 40 C 10 to 95 Non condensing 82 4 F 28 C Maximu...

Page 8: ...w at the last Continuous Flow setting After another 15 seconds the Eclipse 5 stops delivering Continuous Flow and waits 15 seconds for a breath The Eclipse 5 will stay in this modality until a breath...

Page 9: ...gnetic Compatibility Emissions EN55011 FCC Part J15 Class B Conducted EN55011 FCC Part J15 Class B Radiated EN61000 3 2 EN61000 3 3 EN61000 4 2 EN61000 4 3 EN61000 4 4 EN61000 4 5 EN61000 4 6 EN61000...

Page 10: ...provided in this manual Portable and mobile RF communications equipment can affect Medical Electrical Equipment The use of Accessories transducers and cables other than those specified with the excep...

Page 11: ...ons it is recommended that the Eclipse 5 is powered from an uninterruptible power supply UPS Power frequency magnetic field IEC 61000 4 8 30A m Power frequency magnetic fields should be at levels char...

Page 12: ...nd controls both the power source and the Power Cartridge charger Figure 1 Eclipse 5 System Block Diagram ATF CONCENTRATOR MODULE The Eclipse 5 uses a passive system to separate oxygen from air Air fl...

Page 13: ...sor enclosure serves to efficiently draw cooling air in over the compressor cylinders while simultaneously pushing exhaust gas out of the concentrator POWER DISTRIBUTION The Power Manager takes extern...

Page 14: ...ng has been saved and will be used at start up Power Cartridge battery Status Gauge This indicator displays the charge remaining in the battery Each of the five horizontal gray bars represents approxi...

Page 15: ...y rates Regardless of setting the pulse profile is a simple square wave pulse based on a 16 mL volume Refer to Pulse Profiles table below Pulse durations are no less than 100 milli seconds at the begi...

Page 16: ...S x PS 1 is the most sensitive setting PS 3 is the least sensitive setting Factory default is PS 2 HOURS OF OPERATION The Eclipse 5 hour meter provides valuable information on hours of operation Provi...

Page 17: ...t needs EDAT software is capable of sensing and displaying historical event logs and values of components and sensors within the Eclipse 5 system This can be used to determine system faults or user er...

Page 18: ...t is illuminated when the AC Power Supply is supplying 28VDC power If the LED is not illuminated there is no input power available In addition the Power Supply contains protection circuits for output...

Page 19: ...f the user If the Eclipse 5 power cartridge is being charged the battery symbol will display with a waterfall animation However if the power cartridge is not charging then the battery symbol will rema...

Page 20: ...d with the Eclipse 5 is not fully charged Before using the Eclipse 5 Oxygen System for the first time the Power Cartridge must be fully charged Store the power cartridge in a cool dry location Do not...

Reviews: