background image

Oxygen Concentrator 

Technical Service Manual 

| PN 20631679 Rev G       

13

 

 

 

 

COMPRESSOR AND COMPRESSOR ENCLOSURE

 

The Eclipse 5 Compressor is a two-cylinder, variable speed wobble piston compressor, driven by a highly efficient Brushless 
DC (BLDC) motor. When air flows into the Compressor enclosure, it passes through an air intake filter/muffler that muffles 
sound and filters out impurities. Using one cylinder, the compressor takes in filtered air and delivers it to the ATF Module 
under pressure. The second cylinder draws a vacuum on the ATF module and exhausts nitrogen rich gas to the exhaust vent.

 

 

Using  a  multifaceted  approach,  sound,  heat,  and  vibration  generated  by  the  compressor  are  mitigated  by  the  compressor 
enclosure. Vibration and structure-borne noise are addressed by the dual axis gimbal that supports the compressor and the 
tubing that connects the compressor to the ATF module. The rigid walls of the compressor enclosure and the sound adsorbing 
foam that lines it diminish  the  radiated noise.  The centrifugal blower mounted within  the  compressor enclosure serves to 
efficiently draw cooling air in over the compressor cylinders while simultaneously pushing exhaust gas out of the concentrator.

 

 

POWER DISTRIBUTION

 

The Power  Manager  takes  external power  that  comes into  the Eclipse 5 from the power  supplies or Power  Cartridge  and 
monitors and controls power distribution to the rest of the Eclipse 5 system. The Power Manager drives the compressor, ATF 
module motor, blower, and provides power to the Control Board. In addition, when the unit is connected to an external power 
source, the power manager monitors and controls the recharging of the Power Cartridge.

 

CONTROL BOARD

 

The Control Board is at the center of nearly all Eclipse 5 functions. The board constantly monitors dynamics such as temperatures, 
pressures, product flow and concentration, and user input. It determines proper compressor and ATF motor speeds needed in 
order to provide optimum system performance. In addition, this system supports the operation of the Control Panel and its 
indicators.

 

 

The Control Board utilizes a proprietary ultrasonic flow and concentration sensor and a flow control valve to accurately control 
the flow of oxygen in Continuous Flow and Pulse Dose Modes.

 

 
 
 
 

 

Figure 2: Flow Sensor Board Diagram

 

Product Pressure

 

Control Board

 

Summary of Contents for Eclipse 5 with autoSAT

Page 1: ...Eclipse 5 with autoSAT Oxygen Concentrator TECHNICAL SERVICE MANUAL...

Page 2: ...AC Power Supply 18 DC Power Supply 18 Power Cartridge 19 Typical New Power Cartridge Operation Time 20 Charging Algorithm 22 Training The Patient 23 Introduction 23 Pre Delivery Check List 23 Indicati...

Page 3: ...ication Tests 48 Record Hours of Operation Software Version 49 Electrical Safety Test 50 Cleaning the Eclipse 5 50 Provider Service and Maintenance Record 51 Shipping and Transporting the Eclipse 5 52...

Page 4: ...rinciples of operation Providers are given proper guidance in the use of the Eclipse 5 and its accessories that can beconveyed to patients Providers are made aware of the care diagnostics maintenance...

Page 5: ...rd commercial aircraft Flow Setting Indicator Pulse Mode Operation Device operating normally power button Increase Flow Setting Decrease Flow Setting ISO 7000 Graphical symbols for use on equipment In...

Page 6: ...RIDGE BATTERY PN 7082 SEQ PN 20852326S with IEC type A cord 20852325S with IEC type C cord and 20852324S with IEC type G cord two LED model Oxygen Outlet Port Control Panel FAA Approval Icon Rating La...

Page 7: ...0 LPM Continuous Flow 3 0 Pulse Dose Mode 48 dB A 40 dB A Less than 57 dB while alarming Operating Environment Temperature Humidity 50 F to 104 F 10 to 40 C 10 to 95 Non condensing 82 4 F 28 C Maximu...

Page 8: ...w at the last Continuous Flow setting After another 15 seconds the Eclipse 5 stops delivering Continuous Flow and waits 15 seconds for a breath The Eclipse 5 will stay in this modality until a breath...

Page 9: ...gnetic Compatibility Emissions EN55011 FCC Part J15 Class B Conducted EN55011 FCC Part J15 Class B Radiated EN61000 3 2 EN61000 3 3 EN61000 4 2 EN61000 4 3 EN61000 4 4 EN61000 4 5 EN61000 4 6 EN61000...

Page 10: ...provided in this manual Portable and mobile RF communications equipment can affect Medical Electrical Equipment The use of Accessories transducers and cables other than those specified with the excep...

Page 11: ...ons it is recommended that the Eclipse 5 is powered from an uninterruptible power supply UPS Power frequency magnetic field IEC 61000 4 8 30A m Power frequency magnetic fields should be at levels char...

Page 12: ...nd controls both the power source and the Power Cartridge charger Figure 1 Eclipse 5 System Block Diagram ATF CONCENTRATOR MODULE The Eclipse 5 uses a passive system to separate oxygen from air Air fl...

Page 13: ...sor enclosure serves to efficiently draw cooling air in over the compressor cylinders while simultaneously pushing exhaust gas out of the concentrator POWER DISTRIBUTION The Power Manager takes extern...

Page 14: ...ng has been saved and will be used at start up Power Cartridge battery Status Gauge This indicator displays the charge remaining in the battery Each of the five horizontal gray bars represents approxi...

Page 15: ...y rates Regardless of setting the pulse profile is a simple square wave pulse based on a 16 mL volume Refer to Pulse Profiles table below Pulse durations are no less than 100 milli seconds at the begi...

Page 16: ...S x PS 1 is the most sensitive setting PS 3 is the least sensitive setting Factory default is PS 2 HOURS OF OPERATION The Eclipse 5 hour meter provides valuable information on hours of operation Provi...

Page 17: ...t needs EDAT software is capable of sensing and displaying historical event logs and values of components and sensors within the Eclipse 5 system This can be used to determine system faults or user er...

Page 18: ...t is illuminated when the AC Power Supply is supplying 28VDC power If the LED is not illuminated there is no input power available In addition the Power Supply contains protection circuits for output...

Page 19: ...f the user If the Eclipse 5 power cartridge is being charged the battery symbol will display with a waterfall animation However if the power cartridge is not charging then the battery symbol will rema...

Page 20: ...d with the Eclipse 5 is not fully charged Before using the Eclipse 5 Oxygen System for the first time the Power Cartridge must be fully charged Store the power cartridge in a cool dry location Do not...

Reviews: