Oxygen Concentrator
18
Clean and Care for Tubing and
Cannula
Provide your patient instructions on cleaning, disinfection
and/or replacement information for the tubing and cannula.
Clean the Cabinet
To clean the cabinet do the following:
1. Turn OFF the Companion 5 and disconnect from AC
power before any cleaning or disinfection activity.
2. Use mild detergent and water solution.
3. Use a damp (not soaking wet) cloth or sponge.
4. Spray or wet the cloth or sponge with the mild detergent
solution. DO NOT spray the cabinet.
5. Wipe down the cabinet.
6. To disinfect the Companion 5, use Lysol® Brand II
disinfec-tant (or equivalent). Spray or wet a cloth or
sponge with the disinfectant. DO NOT spray the cabinet
or the LED/LCD display. Proceed as directed by the
manufacturer.
Patient Training Checklist
Use the following checklist as a guide to assist in setup and training a patient on the use of the Companion 5 and its accessories.
Patient Name:
Companion 5 Serial #:
Training Topic
Initials
Pre-Delivery Check List
Indications for Use
Contraindications
Basic Concept Training
Advise to read the Users Manual
Safety Guidelines and Operational Safety Warnings/Cautions
Locating the Companion 5
Indicators
Alerts and Alarms
Companion 5 Maintenance
Clean and Care for the Cannula per manufacturer’s instructions.
Clean the Cabinet as needed.
Schedule Intake Filter Replacement and bottom gross particle filter replacement every 2 Years.
Cleaning of optional handle gross particle filter weekly (if present).
Trained By:
Date: